Device recalls

Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N

Medline Industries, LP · The kits contain a Retrograde Coronary Sinus Perfusion (RCSP) cannulae that has the potential for a sterile barrier breach. Sterility of the product cannot be…

Open, ClassifiedPackaging process controlZ-3137-2026
Open FDA's page for this record
Cfres ID
221907
Product res number
Z-3137-2026
Event date initiated
2026-07-23
Event date posted
2026-09-11
Recall status
Open, Classified
Res event number
99583
Product code
LRO
Product description
Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N
Code info
UDI-DI: 10198459445941 (each), 40198459445942 (case); Lot Numbers: 25HLA462, 25HLA916, 25JLA847, 25LLA298, 26CLA372
Firm FEI number
1417592
Recalling firm
Medline Industries, LP
Address 1
3 Lakes Dr
City
Northfield
State
IL
Postal code
60093-2753
Additional info contact
Haley Barclay 866-359-1704
Root cause description
Packaging process control
Product quantity
135 units
Distribution pattern
US Nationwide distribution in the state of CA.
Reason for recall

The kits contain a Retrograde Coronary Sinus Perfusion (RCSP) cannulae that has the potential for a sterile barrier breach. Sterility of the product cannot be assured. If a compromised pouch seal is not identified prior to use and the cannula is used, there is a potential risk of infection or other complications associated with the use of a non-sterile device, including hemolysis, foreign body reaction, thromboembolism, and/or organ dysfunction.

Action

An URGENT MEDICAL DEVICE RECALL notification letter dated 7/23/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-121-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.

K numbers
K213481
Harmonized identifiers (openFDA)
Device name
General Surgery Tray
Medical specialty description
General, Plastic Surgery
Regulation number
878.4370
Device class
2
K number
K854095, K926269, K854292, K962826, K941846, K945160, K963721, K030721, K890570, K935483, K895202, K903336, K894685, K932221, K902887, K980137, K871873, K935740, K925467, K884093, K951149, K023276, K970187, K871322, K901856, K871434, K890098, K905459, K885090, K951478, K902888, K871437, K041080, K905004, K932220, K935743, K910365, K935686, K892389, K925016 +70 more
Registration number
1417592, 3004361903, 1526711, 2431166, 3014704068, 3005997949, 2029275, 3004111573, 3017430544, 3011494917, 3005669815, 3019798849, 1423395, 3006846316, 3004122598, 3020983482, 1000124971, 1055892, 2024980, 3004513970, 3018116884, 2000028362, 3042480961, 3007344957, 3005400762, 3006891803, 1643817, 3005246893, 1527607, 3004605321, 1063382, 1644019, 2936485, 3023140882, 3005012805, 3006973666, 3013466775, 3007139365, 1650907, 1824619 +122 more
FEI number
1417592, 3004361903, 3003803331, 3002692512, 1526711, 1000121355, 3014704068, 1000514656, 3005997949, 3017430544, 3004111573, 3019798849, 3011494917, 3005669815, 3006846316, 3008875470, 3004122598, 3020983482, 1000124971, 3004513970, 3018116884, 2000028362, 3042480961, 3007344957, 3005400762, 3006891803, 1643817, 1000123297, 3005246893, 1527607, 3004605321, 2936485, 3023140882, 3003807323, 3005012805, 3006973666, 1000112877, 3013466775, 3007139365, 1000160458 +122 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.