PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353

GE Medical Systems Information Technologies Inc · Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults fol…

Open, ClassifiedDevice DesignZ-3038-2026
Open FDA's page for this record
Cfres ID
221791
Product res number
Z-3038-2026
Event date initiated
2026-07-20
Event date posted
2026-08-27
Recall status
Open, Classified
Res event number
99536
Product code
MWI
Product description
PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353
Firm FEI number
3001570235
Recalling firm
GE Medical Systems Information Technologies Inc
Address 1
3114 N Grandview Blvd
City
Waukesha
State
WI
Postal code
53188-1615
Additional info contact
GE HealthCare Service 800-437-1171
Root cause description
Device Design
Product quantity
15 units
Distribution pattern
US: MI, NY and Australia, Czech Republic, Finland, Germany, Italy, Netherlands, Qatar, Saudi Arabia, Switzerland, United Arab Emirates, United Kingdom.
Code info

GTIN 00195278364838, Serial Numbers: SUC23080050TA, SUD22470001TA, SUD22420001TA, SUD22470010TA, SUD23390001TA, SUD23390003TA, SUD23390004TA, SUD23390009TA, SUD23390010TA, SUD22470009TA, SUD22470003TA, SUD22470004TA, SUD22470005TA, SUD23390005TA, SUD22280002TA.

Reason for recall

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.

Action

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 7/20/2026 by letter delivered using traceable means. The notice explained the issue, potential risk, and requested the following: "Actions to be taken by Customer/User You can continue to use your Portrait Mobile Monitoring Solution with Portrait Core Services SW V1.0.5, V1.1 or V1.2. Please ensure appropriate alternative monitoring is established for patients before a software update is started. A User Manual addendum is included with this letter that includes specific instructions related to alternative monitoring. Please place a copy of the addendum in Appendix A with the User Manual for your Portrait Mobile Monitoring Solution with Portrait Core Services SW V1.0.5, V1.1 or V1.2 system in your facility. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

K numbers
K230626
Harmonized identifiers (openFDA)
Device name
Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
Medical specialty description
Cardiovascular
Regulation number
870.2300
Device class
2
K number
K080173, K053112, K151839, K053193, K150869, K210204, K151644, K113165, K143751, K121013, K123711, K183370, K013019, K101571, K093017, K181919, K152330, K153225, K131898, K082256, K073094, K170032, K134014, K170168, K122133, K013470, K010732, K133450, K003622, K101127, K143372, K172200, K071802, K072516, K130740, K153739, K010342, K000404, K063419, K170047 +268 more
Registration number
3038259592, 1316463, 3011683100, 8022257, 3007137598, 1221108, 3004091281, 3021632375, 3006606827, 3012238587, 3010097171, 3006648320, 3008977358, 3011353843, 3001570235, 3027256461, 3009673645, 3019834104, 3034603598, 3016670256, 9610105, 9680215, 3032013079, 8020045, 3004415095, 1222630, 3008787503, 1833920, 3023412068, 3007836437, 3003174284, 3035111606, 3009001657, 3033959233, 3007481893, 3013596024, 2520313, 3021219789, 3007029079, 3042175844 +136 more
FEI number
3038259592, 1316463, 3011683100, 3004117725, 3004631140, 3007137598, 1221108, 3006606827, 3021632375, 1000292048, 3012238587, 3010097171, 3006648320, 3008977358, 3011353843, 3001570235, 3002807880, 3009673645, 3019834104, 3027256461, 3034603598, 3002807785, 3016670256, 3032013079, 3004415095, 3008787503, 3023412068, 3007836437, 3000957797, 3005154439, 3003174284, 3035111606, 3009001657, 3033959233, 3007481893, 3013596024, 2520313, 3021219789, 3007029079, 3042175844 +136 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.