Optima XR220amx, Mobile Digital-Ready Radiographic System

GE Medical Systems, LLC · GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety p…

Open, ClassifiedProcess controlZ-2727-2026
Open FDA's page for this record
Cfres ID
221185
Product res number
Z-2727-2026
Event date initiated
2026-06-05
Event date posted
2026-07-13
Recall status
Open, Classified
Res event number
99279
Product code
IZL
Product description
Optima XR220amx, Mobile Digital-Ready Radiographic System
Firm FEI number
2126677
Recalling firm
GE Medical Systems, LLC
Address 1
3000 N Grandview Blvd
City
Waukesha
State
WI
Postal code
53188-1615
Additional info contact
GE HealthCare Service 800-437-1171
Root cause description
Process control
Product quantity
38 units
Distribution pattern
Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.
Code info

UDI-DI: 00840682115261; System IDs: 05249MOB02.OLD, 205820DP1, 228818XR220, 229276M1, 240566P1, 301295OPTIMA2, 303788SBOPT1, 314251P5, 352LEEDP1, 386FLGDPXR3, 417347DPORT1, 512SAMCSB, 516632OPTIMA2, 540981ODP7, 563264OPT, 601445DP1, 615338HDP01, 661949DP2, 713441OPCDX1, 717657OPT1, 718206OPT240A, 727848SB1, 732321DP1, 757736DP3, 813302OPT1, 828213DXP1, 828286OPT1, 843847220D, 850070941, 865545DP3, 915821OPT1, 941342OP221, 972566XR220E, 985447TM4, SA1208MM05, SB2201700085WK, XR1034798WK7, XR1037005WK4

Reason for recall

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

Action

An URGENT MEDICAL DEVICE CORRECTION notification letter dated 6/5/26 was sent to customers. Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use the system by following the instructions below: Perform a check of your device prior to each patient exam to look for the following behaviors: "Column vertical movement requires increased effort. "Column has change in resistance (i.e. movement is not smooth) during vertical movement. "Column makes abnormal noises during vertical movement. If any are observed, DO NOT use the device and call GE HealthCare Service Immediately. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 10967 Digital Response Form or print, fill out manually, scan, and email to Recall.10967@gehealthcare.com.

K numbers
K142383, K103476
Harmonized identifiers (openFDA)
Device name
System, X-Ray, Mobile
Medical specialty description
Radiology
Regulation number
892.1720
Device class
2
K number
K220700, K020436, K244049, K790392, K251223, K200637, K182639, K222896, K924798, K770903, K140723, K211473, K945043, K212654, K222339, K251443, K891143, K002685, K203782, K052721, K833226, K200726, K221417, K192011, K931483, K191813, K070634, K932106, K232185, K213462, K151038, K171119, K781676, K173602, K201488, K192936, K234108, K955095, K122697, K203658 +198 more
Registration number
3013779985, 3005810333, 3033527770, 3012127758, 1418964, 3007836437, 3009717300, 3017375271, 3025442726, 3008496839, 3011015597, 3008044777, 3024891354, 2433615, 3023003446, 3005877899, 3013677815, 3010398867, 3001722928, 3009715136, 3014809646, 3003775006, 3015276088, 8030233, 3007065358, 3031503890, 9615439, 1418484, 3032116890, 9614430, 3018269927, 3030089787, 3018077344, 3004063527, 3016250464, 3004706313, 3003768251, 3016674851, 3008284470, 3009077524 +39 more
FEI number
3013779985, 3005810333, 3033527770, 3012127758, 1418964, 3007836437, 3002495042, 3002808139, 3009717300, 1000120772, 3017375271, 3025442726, 3002143662, 3008496839, 3011015597, 3008044777, 3024891354, 2433615, 3023003446, 3005877899, 3013677815, 3010398867, 3003377038, 3001722928, 3009715136, 3014809646, 3003775006, 3015276088, 3007065358, 3031503890, 1418484, 3032116890, 3018269927, 3030089787, 3018077344, 3004063527, 3016250464, 3004706313, 3003768251, 3016674851 +39 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.