Device recalls

IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized tumors…

Ion Beam Applications S.A. · Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances

Open, ClassifiedDevice DesignZ-1945-2024
Open FDA's page for this record
Cfres ID
207610
Product res number
Z-1945-2024
Event date initiated
2024-05-10
Event date posted
2024-06-03
Recall status
Open, Classified
Res event number
94576
Product code
LHN
Firm FEI number
3000256071
Recalling firm
Ion Beam Applications S.A.
Address 1
Chemin Du Cyclotron 3 3 chemin du cyclotron
City
Ottignies-Louvain-La-Neuve
Country
Belgium
Additional info contact
SAME 104758811
Reason for recall
Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances
Root cause description
Device Design
Product quantity
2 units
Distribution pattern
Worldwide - US Nationwide distribution in the state of Virginia and the country of India.
Product description

IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Code info

UDI: SAT140: (01)05404013814015(11)190507(21)SAT140 SAT136: (01)05404013801138(11)201105(21)SAT136 Serial Numbers: SAT136, SAT140 PTS-11 versions from PTS-11.1 and PTS-12 versions, with C230 accelerator

Action

Iba issued Urgent Medical Device Correction Letter on May 10, 2024 via IBA responsible directly to customer site either by hand or by email. Letter states reason for recall, health risk and action to take: ACTION BEING TAKEN BY IBA Monitoring of the deflector, software upgrade. 7.2 CORRECTIVE ACTIONS Software upgrade 7.3 PREVENTIVE ACTIONS Waiting for the solution to be deployed, IBA will distribute an Internal User Notice to IBA operators of impacted sites with instructions to monitor the deflector and stop the irradiation when necessary. 7.4 TIME SCHEDULE FOR THE IMPLEMENTATION OF THE DIFFERENT ACTIONS The Internal User Notice will be provided to IBA operators by May 10, 2024. The corrective action will be implemented at your site by June 2025. Contact: Sonia PINEL Vigilance@iba-group.com

K numbers
K163500
Harmonized identifiers (openFDA)
Device name
System, Radiation Therapy, Charged-Particle, Medical
Medical specialty description
Radiology
Regulation number
892.5050
Device class
2
K number
K891261, K132919, K162246, K170719, K160612, K161163, K100766, K151132, K232032, K983342, K133191, K130426, K972212, K112768, K071399, K220883, K852907, K132847, K163500, K120676, K181789, K160063, K260174, K231863, K133914, K200282, K191821, K061913, K191521, K082416, K032511, K992414, K983332, K972211, K062891, K092796, K181498, K851301, K002194, K061051 +53 more
Registration number
2247992, 8043933, 2916710, 3006579682, 3009754495, 9613183, 9681124, 1062572, 3010658895, 3010726836, 2953704, 3003993895, 1721113, 2032229, 3000256071, 3005374502, 3007087027, 9614952, 3015178562, 3004977335, 1000162206, 3008317830, 3015232217, 3016872563, 3042022819, 1038814, 8031557
FEI number
3006653576, 2916710, 3006579682, 3009754495, 3007360740, 3003166990, 3010658895, 3010726836, 3002619595, 1270485, 1000227792, 3003993895, 1721113, 3000256071, 3007087027, 3015178562, 3001238439, 1000208827, 3004977335, 1000162206, 3001874253, 3008328959, 3015232217, 3016872563, 3042022819, 1038814, 3008317830