Trevo Pro 14 Microcatheter, Catalog: 90231; Trevo Pro 18 Microcatheter, Catalog: 90238
Stryker Neurovascular · Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAM…
Open FDA's page for this record- Cfres ID
- 207236
- Product res number
- Z-1904-2024
- Event date initiated
- 2024-04-23
- Event date posted
- 2024-05-28
- Recall status
- Open, Classified
- Res event number
- 94437
- Product code
- DQO
- Product description
- Trevo Pro 14 Microcatheter, Catalog: 90231; Trevo Pro 18 Microcatheter, Catalog: 90238
- Code info
- Catalog/UDI-DI/Lot: 90231/00815742002317/121050, 129849, 129882, 130285, 136439, 136628, 143375, 143476; 90238/00815742002386/136009, 198334
- Firm FEI number
- 3008853977
- Recalling firm
- Stryker Neurovascular
- Address 1
- 47900 Bayside Pkwy
- City
- Fremont
- State
- CA
- Postal code
- 94538-6515
- Root cause description
- Process design
- Product quantity
- 648
- Reason for recall
Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
- Action
On 4/23/24, recall notices were mailed to Supply Chain Management, Recall Coordinators, and Inventory Managers who were asked to do the following: 1) Cease further distribution or use of all affected devices. 2) Segregate the affected units in a secure location for return to the firm. 3) Circulate this Recall-Removal notice internally to all interested/affected parties. 4) Maintain awareness of this communication internally until all required actions have been completed within your facility. 5) Inform the firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so the firm can inform the recipients appropriately. 6) Inform the firm of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916-3876), Email: NVCustomerCare@stryker.com) 7) Complete and return the response form via email to nvfieldactions@stryker.com
- Distribution pattern
Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.
- K numbers
- K131437, K113260
- Harmonized identifiers (openFDA)
- Device name
- Catheter, Intravascular, Diagnostic
- Medical specialty description
- Cardiovascular
- Regulation number
- 870.1200
- Device class
- 2
- K number
- K944284, K914009, K885348, K974559, K790107, K052637, K911703, K823758, K060116, K860429, K801321, K230620, K930479, K212977, K993694, K041291, K062627, K911270, K150232, K771810, K813623, K830523, K133130, K942939, K171264, K981441, K024126, K770290, K121694, K193647, K830788, K853182, K173762, K012983, K915726, K854018, K131770, K841615, K041909, K884354 +461 more
- Registration number
- 8020616, 3003678543, 3013351723, 3004111573, 2024168, 1646831, 3014656749, 1721686, 3007830707, 3006425876, 3014779787, 2183744, 3011642792, 1319211, 2029214, 3008264254, 3006950086, 3015225571, 3004122598, 8010026, 9616683, 3012494290, 1928237, 1055236, 3005669815, 2521402, 9616099, 3008496528, 3011660785, 3004605321, 1058584, 3012497308, 1061927, 9612152, 2528981, 3007289408, 3005941719, 3021343210, 1123137, 9618000 +171 more
- FEI number
- 3003678543, 3013351723, 3002806476, 3004111573, 2024168, 3014656749, 1721686, 3004970277, 3007830707, 3006425876, 3014779787, 2183744, 3011642792, 1319211, 1000514656, 3003807466, 1000306431, 3008264254, 3006950086, 3015225571, 3004122598, 3003040625, 3012494290, 1928237, 1000112877, 3005669815, 3016261803, 2521402, 3004605321, 3008496528, 3011660785, 3002534070, 3003094851, 3012497308, 1000512168, 3007289408, 2528981, 3002807548, 3021343210, 3004467263 +170 more