Device recalls

Trevo XP ProVue Retriever 4 x 20, Catalog: 90182 Trevo XP ProVue Retriever 3 x 20, Catalog: 90183 Trevo XP ProVue Retriever 4 x 30, Catalog…

Stryker Neurovascular · Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAM…

Open, ClassifiedProcess designZ-1902-2024
Open FDA's page for this record
Cfres ID
207220
Product res number
Z-1902-2024
Event date initiated
2024-04-23
Event date posted
2024-05-28
Recall status
Open, Classified
Res event number
94437
Product code
POL
Firm FEI number
3008853977
Recalling firm
Stryker Neurovascular
Address 1
47900 Bayside Pkwy
City
Fremont
State
CA
Postal code
94538-6515
Root cause description
Process design
Product quantity
461
Product description

Trevo XP ProVue Retriever 4 x 20, Catalog: 90182 Trevo XP ProVue Retriever 3 x 20, Catalog: 90183 Trevo XP ProVue Retriever 4 x 30, Catalog: 90185 Trevo XP ProVue Retriever 6 x 25, Catalog: 90186 Kits: TREVO XP 3X20+PRO14 2-PACK, Catalog: 80051 TREVO XP 4X20+PRO18 2-PACK, Catalog: 80052 TREVO XP 4X30+PRO18 2-PACK, Catalog: 93067 TREVO XP 6X25+XT27 2-PACK, Catalog: 93068 TREVO XP 3X20+PRO14+CAT6 3-PACK - CE, Catalog: M0033PK32021002 TREVO XP 4X20+PRO18+CAT6 3-PACK - CE, Catalog: M0033PK42022002 TREVO XP 6X25+XT27+CAT6 3-PACK - CE, Catalog: M0033PK62523002

Code info

Catalog/UDI-DI/Lot: 90182/00815742001822/116706; 90183/00815742001839/117514; 90185/00815742001853/118900; 90186/00815742001860/114054; 80051/00815742000511/47307561, 47328877, 47351811, 47398058, 47398074, 48521577, 48521585, 48540280, 48540281, 48540282, 48540283, 48540284, 48556786, 48588490, 48607589, 48676218, 48696629, 48715720, 48736116, 48736122, 48754915, 48754920, 48772012, 48772014, 48772025, 50065499, 50065511, 50065512, 50065513, 50065514, 50065515, 50065518, 50113384, 50113385, 50133176, 50133191, 50193582, 50216273, 50216274, 50216277, 50239676, 50260575, 50260583, 50280979, 50281011, 50304688, 50304712, 50304713, 50304714, 50397877, 50422572, 50422582, 50422589, 50448279, 50519274, 50519275, 50519276, 50519277, 50519285, 50519295, 50577572, 50607374, 50607376, 50607377, 50607378, 50650278, 50650279, 50650288, 50673881, 50673884, 50673908, 50673909, 50673910, 50673911, 50673930, 50694279, 50738912, 50786710, 50850697, 50869887, 50869892, 50869905, 50886487, 50912493, 50912514, 50912515, 50912516, 50912517, 50931490, 50931518, 50931519, 50950402, 50967093, 51320497, 51350503, 51350505, 51489228, 51489235, 51513311, 51513516, 51535421, 51729133; 80052/00815742000528/49021114, 49021115, 49021116, 49021117, 49021118, 49033114, 49033115, 49033116, 49033117, 49052517, 49052525, 49192125, 49192140, 49192141, 49216723, 49216724, 49216737, 53175402; 93067/00815742000672/53320201; 93068/00815742000689/48963313, 49056315, 49155966, 49216725, 49318831, 49318836, 49318837, 49418740, 49418743, 49441346, 50000932, 50000933, 50000934, 50029333, 50029340, 50029361, 50047054, 50065482, 50065485, 50065486, 50065487; M0033PK32021002/07613327126334/48500990, 48624226, 48641523, 48641533, 50156372, 50519272, 50650290, 50912513, 50967091; M0033PK42022002/07613327126358/48995917, 49033123, 49140624, 49236926; M0033PK62523002/07613327126365/49033118, 49236935, 50047053

Reason for recall

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

Action

On 4/23/24, recall notices were mailed to Supply Chain Management, Recall Coordinators, and Inventory Managers who were asked to do the following: 1) Cease further distribution or use of all affected devices. 2) Segregate the affected units in a secure location for return to the firm. 3) Circulate this Recall-Removal notice internally to all interested/affected parties. 4) Maintain awareness of this communication internally until all required actions have been completed within your facility. 5) Inform the firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so the firm can inform the recipients appropriately. 6) Inform the firm of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916-3876), Email: NVCustomerCare@stryker.com) 7) Complete and return the response form via email to nvfieldactions@stryker.com

Distribution pattern

Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.

K numbers
K173352
Harmonized identifiers (openFDA)
Device name
Neurovascular Mechanical Thrombectomy Device For Acute Ischemic Stroke Treatment
Medical specialty description
Neurology
Regulation number
882.5600
Device class
2
K number
K190779, K223305, K222848, K200117, DEN150049, K231539, K193576, K183022, K203219, K162539, K181060, K203358, K242676, K192207, K173352, K181807, K181186, K210502
Registration number
3013656790, 3008853977, 2029214, 3031288142, 3015550451, 3016540856, 3007007790, 2011171, 3006451981, 3012931345, 3012497308, 1721676, 3009380063
FEI number
3013656790, 3008853977, 3031288142, 3015550451, 3016540856, 3007007790, 2011171, 3006451981, 3012931345, 1000220862, 3012497308, 1721676, 3009380063
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.