FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremi…

FUJIFILM Healthcare Americas Corporation · Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury

Open, ClassifiedComponent design/selectionZ-1898-2024
Open FDA's page for this record
Cfres ID
207224
Product res number
Z-1898-2024
Event date initiated
2024-04-11
Event date posted
2024-05-24
Recall status
Open, Classified
Res event number
94443
Product code
KPR
Firm FEI number
1000513161
Recalling firm
FUJIFILM Healthcare Americas Corporation
Address 1
81 Hartwell Ave Ste 300 81 Hartwell Ave
Address 2
Ste 300
City
Lexington
State
MA
Postal code
02421-3160
Additional info contact
SAME 617-861-5300
Reason for recall
Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury
Root cause description
Component design/selection
Product quantity
18 units
Distribution pattern
US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.
Product description

FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. Model/Catalog Number: 566-16130-23 566-16130-33

Code info

UDI-DI: 04540217052226 04540217057450 S/N: MP95A9F6A001, MP95A9F74002, MP95A9F74001, MP95AA983002, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA993001, MP95AA996002, MP95AA996001, MP95AA997001, MP95A9F6B001, MP95A9F76001, MP95AA983001, MP95AA988002, MP95AA98A001, MP95AA993002

Action

FUJIFILM Healthcare Americas Corporation (FUJIFILM) issued Urgent -FIrst Notice letter via email on 4/11/24. Letter states reason for recall, health risk and action to take: 1)Please complete and return the form attached on page 3 of this communication. Providingthe response with the information requested is essential for ensuring appropriate action istaken 2)Please contact Customer Service at 1-888-FUJIMED (1-888-385-4633) to open a serviceticket. 3)Until your upgrade is scheduled and completed, you can continue to safely use FDRVisionary Suite, however, inspect the bolts prior to use. Please contact our field action coordinator, Chaitrali Kulkarni, via email at hcusregulatoryaffairs@fujifilm.com, if you have any further questions regarding this field action.

K numbers
K152294
Harmonized identifiers (openFDA)
Device name
System, X-Ray, Stationary
Medical specialty description
Radiology
Regulation number
892.1680
Device class
2
K number
K102680, K082064, K152788, K904145, K163203, K150506, K883160, K050704, K152767, K223550, K982955, K073632, K934005, K233722, K962896, K940306, K790741, K073114, K122454, K173517, K920347, K231892, K052541, K071430, K111655, K193637, K242119, K220919, K852349, K113855, K191699, K902791, K203340, K981880, K090238, K192541, K080655, K915362, K800788, K050151 +319 more
Registration number
3021059265, 9616014, 3009529630, 3001556265, 3024891354, 9614430, 2126677, 3003202425, 3010300212, 3001722928, 2243057, 3003768251, 3008284470, 3016674851, 8030233, 3013679558, 3016250464, 3013714596, 3029916061, 3006115340, 3009134655, 1450503, 3006051164, 3003768277, 3014209556, 3033525986, 3004977335, 9616047, 3009608361, 3033527770, 3011416394, 3008364043, 3025442726, 1317307, 3027339477, 3005810333, 3030089787, 3006165260, 3012569470, 3035950510 +43 more
FEI number
3021059265, 1000114111, 3002808139, 3009529630, 3001556265, 3024891354, 2126677, 3003202425, 3010300212, 3001722928, 2243057, 3003768251, 3008284470, 3016674851, 3013679558, 3016250464, 3002495042, 3013714596, 3029916061, 3006115340, 3009134655, 1450503, 3006051164, 3003377038, 3014209556, 3033525986, 3004977335, 3009608361, 3033527770, 3011416394, 3008364043, 3025442726, 1317307, 3027339477, 3002175190, 3005810333, 3030089787, 3002143662, 3006165260, 3012569470 +43 more