8000-021-002, Scopis ENT Software with TGS with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating funct…

Stryker Leibinger GmbH & Co. KG · Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference p…

Open, ClassifiedSoftware Manufacturing/Software DeploymentZ-1885-2024
Open FDA's page for this record
Cfres ID
207243
Product res number
Z-1885-2024
Event date initiated
2024-04-04
Event date posted
2024-05-23
Recall status
Open, Classified
Res event number
94456
Product code
PGW
Firm FEI number
3002821504
Recalling firm
Stryker Leibinger GmbH & Co. KG
Address 1
Botzinger Str. 41 41 botzinger str.
City
Freiburg Im Breisgau
Country
Germany
Additional info contact
Lacey Marshall 269-270-5950 Ext. 0
Root cause description
Software Manufacturing/Software Deployment
Product quantity
705 units
Distribution pattern
US Nationwide distribution.
Product description

8000-021-002, Scopis ENT Software with TGS with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional endoscopic sinus surgery (FESS) offering surgeons highly advanced image guidance and visualization capabilities in a single system. The Stryker Electromagnetic Navigation Unit (8000-010-003) powered by Scopis ENT Software with TGS (8000-021-002), allows the analysis and identification of sinus cells in the complex patient anatomy and planning of the natural drainage pathways through the sinus cavity. During surgery, the planned pathways are overlaid in real-time onto the endoscopic image, providing a unique Scopis augmented reality technology. Guidance of endoscopic instruments may help perform a minimally invasive, accurate and selective surgery.

Code info

Sofware Version: NOVA 3.6.0 RC16, Product Code: 8000-021-002; UDI/DI: 07613327514179; Serial Numbers: 10984, 10110, 10164, 10454, 10762, 10387, 10120, 10252, 10414, 10612, 10408, 10673, 11013, 11032, 10566, 10974, 0310-B, 10326, 10442, 10729, 10681, 10552, 10088, 10476, 10307, 10121, 10371, 0252-B, 10262, 10602, 10613, 10381, 10234, 200774, 10135, 0304-B, 10303, 10257, 10639, 10445, 10684, 10740, 10986, 10413, 10079, 10528, 10565, 10482, 10209, 10574, 10606, 10229, 10214, 10797, 0335-B, 10284, 10465, 10295, 10308, 0282-B, 83B750AD, 0287-B, 10763, 10434, 11025, 10578, 10473, 10596, 0291-B, 10624, 10591, 11092, 10758, 10097, 10294, 0251-B, 10597, 10769, 0309-B, 10765, 0313-B, 0281-B, 10513, 11088, 10057, 10337, 10464, 10621, 10699, 10509, 10421, 10549, 10003, 10499, 10661, 10668, 0292-B, 10416, 0263-B, 10654, 10571, 0329-B, 10934, 10798, 11022, 10610, 10766, 10471, 10437, 10648, 10410, 10403, 10159, 10192, 10372, 10189, 10193, 10642, 10390, 10771, 10202, 10806, 10237, 10296, 0244-B, 0307-B, 10034, 10440, 10704, 10101, 10153, 10098, 10388, 10720, 10943, 10944, 10994, 10695, 10608, 10615, 10054, 10076, 10015, 10089, 10208, 10776, 10780, 11111, 10586, 10759, 10431, 10772, 10376, 10563, 10978, 10989, 10369, 10641, 10796, 10558, 10660, 10071, 10987, 10953, 10982, 10744, 10026, 0325-B, 10959, 11000, 10453, 0332-B, 10055, 10335, 10742, 10649, 10983, 10512, 10958, 10268, 10640, 0280-B, 11075, 11117, 10356, 10429, 10432, 10659, 10677, 10319, 10311, 10332, 10757, 10809, 10523, 10382, 10451, 10495, 10997, 11116, 11018, 10083, 10672, 10775, 10053, 10195, 10330, 10161, 10007, 10328, 10949, 11031, 10781, 10990, 10350, 10750, 10191, 10033, 10056, 10446, 10078, 10516, 10373, 10722, 10136, 10210, 10091, 10255, 10253, 10060, 10427, 10162, 10297, 10731, 10935, 0285-B, 10336, 10760, 10386, 10347, 10286, 10165, 10041, 10456, 10312, 10417, 10500, 10133, 10224, 10031, 10040, 10100, 10784, 10248, 10423, 10215, 10951, 10092, 0260-B, 10185, 10127, 10112, 10790, 10299, 10316, 10435, 10459, 10075, 10479, 10357, 10627, 10732, 10734, 0343-B, 0320-B, 10517, 10455, 10590, 11030, 10383, 10043, 10096, 10789, 10631, 10489, 10791, 10220, 10658, 10348, 10143, 10114, 10625, 10799, 11108, 10103, 10522, 10126, 10066, 10980, 11102, 0318-B, 11005, 10177, 10662, 10166, 10197, 10300, 10406, 10181, 0326-B, 10443, 10331, 10364, 10368, 10592, 10062, 10366, 10349, 10353, 11097, 10155, 10218, 10227, 10230, 10315, 10618, 10021, 10787, 11038, 0271-B, 10365, 10637, 10141, 10525, 10341, 10433, 10696, 0333-B, 10401, 10633, 10211, 10231, 10052, 10794, 10685, 10157, 10213, 10081, 10123, 10090, 10199, 10678, 11020, 10216, 10377, 10424, 10212, 10361, 10059, 10109, 10736, 10344, 10693, 10977, 11004, 10016, 10168, 10171, 10691, 10006, 10458, 10324, 10156, 10190, 10221, 10363, 10475, 0322-B, 0339-B, 10692, 10679, 10082, 10573, 10603, 10267, 10541, 10999, 10169, 10628, 10576, 10345, 10738, 10354, 10238, 0259-B, 10077, 0306-B, 10461, 10172, 10243, 10233, 10327, 0327-B, 11023, 11114, 10450, 10398, 10404, 0303-B, 10400, 10412, 10593, 10072, 10514, 10239, 10069, 10244, 10249, 10761, 10359, 10739, 10352, 10163, 10374, 10698, 10342, 10607, 10635, 10058, 10425, 10419, 10950, 10405, 10422, 10223, 0308-B, 10010, 10594, 10205, 10370, 10474, 10247, 10020, 10174, 10184, 10188, 10128, 11003, 10689, 10701, 10795, 10074, 10266, 10288, 10008, 10203, 10966, 10438, 10955, 10073, 10080, 10333, 10777, 10508, 10325, 10086, 10104, 10469, 10755, 10810, 10671, 10152, 10393, 10657, 10105, 10298, 10524, 10805, 10375, 10402, 10094, 10358, 10498, 10501, 10384, 10391, 10993, 10064, 10175, 10397, 10560, 10570, 10630, 10132, 10579, 10439, 10415, 10306, 10343, 10477, 10323, 10737, 10589, 10304, 10338, 10460, 10598, 10779, 10988, 10346, 11082, 10436, 10362, 10334, 10682, 10653, 10583, 10724, 10725, 10407, 10380, 10409, 10595, 10137, 10656, 10389, 10611, 10638, 11036, 11037, 10577, 10609, 10587, 10626, 10965, 11017, 11112, 10547, 10954, 11029, 11026, 10979, 10985, 11080, 10981, 11099, 10207, 10049, 10580, 10144, 10723.

Reason for recall

Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.

Action

On April 22, 2024 an URGENT MEDICAL DEVICE FIELD ACTION letters were sent to customers. Next Steps: 1. Locate any ENU. Quarantine and discontinue use until software on device has been confirmed. - Locate software version and serial number (see pg. 2 for instructions). - Your Stryker Sales Representatives may help, if needed. 2. Return the enclosed Business Reply Form (BRF) via email to instruments.recalls@stryker.com. - If Scopis ENT Software with TGS version NOVA 3.6.0 RC16 is installed, Stryker will arrange shipment of compliant software. 3. Once compliant software is received and installed, return the enclosed Software Update Form via email to instruments.recalls@stryker.com to confirm completion of the recall action steps. 4. Maintain this notification internally until all required actions have been completed. Please contact Lacey Marshall at 269-270-5950 or Instruments.recalls@stryker.com with questions or concerns. Report any serious adverse events or product quality problems to Stryker Instruments: 1-800-253-3210. Health care professionals and consumers may report serious adverse events or product quality problems with the use of this product to the FDA's MedWatch Adverse Event Reporting program either online, or by fax or phone. Online: ww.fda.gov/Safety/MedWatch/HowToReport/default.htm Fax: (800) FDA-0178 Phone: (800) FDA-1088.

K numbers
K221098
Harmonized identifiers (openFDA)
Device name
Ear, Nose, And Throat Stereotaxic Instrument
Medical specialty description
Neurology
Regulation number
882.4560
Device class
2
K number
K202609, K243053, K160369, K153555, K141551, K193502, K193118, K160479, K180948, K192397, K200041, K262235, K162176, K150473, K201174, K163416, K170018, K190532, K141456, K193453, K133573, K200723, K161555, K211291, K161940, K153247, K173628, K253395, K161701, K261540, K133665, K161491, K161697, K221037, K221098, K231862, K223734, K171661
Registration number
3018094310, 3003678543, 8043933, 3005172759, 3016245399, 1058084, 1811755, 3007738736, 3010273872, 3010202439, 3010313588, 3011874573, 1045254, 1221485, 3006697241, 3018565642, 3009018440, 1928237, 3015967359, 3003500271, 3005677016, 3005423519, 3009217531, 1824206, 3015462848, 1723170, 3003418325, 3010707607, 9680721, 3008837339, 3015131155, 1225492, 3017060084, 3014342096, 8040510, 3008087383, 9610905, 2020550, 8010177, 3010009632 +4 more
FEI number
3018094310, 3003678543, 3005172759, 3001237338, 3016245399, 1811755, 3007738736, 3010273872, 3010313588, 3011874573, 1045254, 3002808268, 1221485, 3006697241, 3018565642, 1000517638, 3002619595, 3009018440, 3002821504, 1000221981, 3001319476, 1928237, 3015967359, 3003500271, 3005677016, 3006165545, 3009217531, 1824206, 3015462848, 3003418325, 3002807464, 3010707607, 3008837339, 3015131155, 3017060084, 3014342096, 3008087383, 2020550, 3007475226, 1000244733 +4 more