Device recalls

Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24

FUJIFILM Healthcare Americas Corporation · AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent in…

Open, ClassifiedDevice DesignZ-1884-2024
Open FDA's page for this record
Cfres ID
207222
Product res number
Z-1884-2024
Event date initiated
2024-04-11
Event date posted
2024-05-22
Recall status
Open, Classified
Res event number
94441
Product code
IYN
Product description
Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24
Code info
UDI-DI: (01) 04580292771704 Noblus Ultrasound Imaging System AC Adapter Serial Number Before K17160001 (Excluding K17160001)
Firm FEI number
1000513161
Recalling firm
FUJIFILM Healthcare Americas Corporation
Address 1
81 Hartwell Ave Ste 300 81 Hartwell Ave
Address 2
Ste 300
City
Lexington
State
MA
Postal code
02421-3160
Additional info contact
SAME 617-861-5300
Root cause description
Device Design
Product quantity
418 units
Distribution pattern
US Nationwide distribution.
Reason for recall

AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).

Action

FujiFilm issued Urgent Medical Device Recall Letter via email on 4/11/24. Letter states reason for recall, health risk and action to take: Please complete and return the Field Action Verification Form attached on page 3 of this communication. Responding with the information requested is essential for ensuring appropriate action is taken. Please accept the replacement of AC adapter that is implemented the countermeasure. Please contact our field action coordinator, Kotei Aoki, via email at HCUSRegulatoryAffairs@fujifilm.com, if you have any further questions regarding this field action.

K numbers
K160559
Harmonized identifiers (openFDA)
Device name
System, Imaging, Pulsed Doppler, Ultrasonic
Medical specialty description
Radiology
Regulation number
892.1550
Device class
2
K number
K121564, K172342, K231190, K173090, K041916, K020630, K110199, K113663, K131088, K121010, K192254, K102388, K070813, K933916, K202364, K173000, K904017, K032620, K233692, K233610, K240206, K042055, K050380, K250543, K180535, K232336, K230179, K233394, K161525, K170823, K000931, K260748, K053648, K122583, K152396, K120321, K231685, K261211, K142710, K051334 +1,149 more
PMA number
P940005
Registration number
3006161756, 3009293490, 3032807464, 3033527241, 3014574557, 3008492462, 3040391009, 9611125, 3010398867, 3043534483, 2518586, 2183744, 3027501622, 3033508292, 3016696283, 1126271, 3008640801, 9615439, 3013368530, 3032109225, 3009196578, 3009758186, 3003851073, 3016842138, 3009247326, 3000976525, 3026047529, 3019784575, 3025429169, 2531814, 3012761544, 9680269, 3042172882, 3010220244, 3012987001, 3017487805, 3001722928, 3047889344, 1058584, 3033557977 +182 more
FEI number
3009293490, 3032807464, 3033527241, 1000114111, 3014574557, 3008492462, 3040391009, 3003861317, 3010398867, 3043534483, 2518586, 2183744, 3027501622, 3002808127, 3033508292, 3016696283, 3008640801, 3003705156, 3013368530, 1000306431, 3032109225, 3000302153, 3009196578, 3009758186, 3003851073, 3016842138, 3009247326, 3000976525, 3026047529, 3019784575, 3025429169, 3012761544, 3042172882, 3010220244, 3012987001, 3017487805, 3001722928, 3047889344, 3033557977, 3008223664 +182 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.