Device recalls

Reagent Test Strips for Blood, Urobilinogen, Bilirubin, Protein, Nitrite, Ketones, Ascorbic Acid, Glucose, pH, Specific Gravity and Leucocy…

Macherey Nagel Gmbh & Co. Kg · The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer…

Open, ClassifiedUnder Investigation by firmZ-1834-2024
Open FDA's page for this record
Cfres ID
207241
Product res number
Z-1834-2024
Event date initiated
2024-04-04
Event date posted
2024-05-16
Recall status
Open, Classified
Res event number
94454
Product code
JIL
Code info
Product #102-2285, UDI/DI: 00304040001479, LOT 65402.
Firm FEI number
4580
Recalling firm
Macherey Nagel Gmbh & Co. Kg
Address 1
Valencienner Str. 11 11 valencienner str.
City
Duren
Country
Germany
Additional info contact
Dr. Markus Meusel 49-2421 969 21
Root cause description
Under Investigation by firm
Product quantity
5,200 packages
Distribution pattern
US Nationwide distribution in the states of CO, FL, IN, NV, TX.
Product description

Reagent Test Strips for Blood, Urobilinogen, Bilirubin, Protein, Nitrite, Ketones, Ascorbic Acid, Glucose, pH, Specific Gravity and Leucocytes in Urine by the Dip and Read Technique.

Reason for recall

The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer by way of a Henry Schein Inc. (HSI) customer complaint. This problem can potentially result in the possibility that the parameter leucocytes (a non-specific marker for urinary track infections and an indicator for inflammatory renal processes) is interpreted instrumentally as a false negative. The following HSI item number and lot number were affected: Urispec 11-Way, #102-2285, LOT 65402.

Action

On April4, 2024 URGENT FIELD SAFETY NOTICE letters were sent to direct consignees instructing them to do the following: For distributors: 1. The product named in this safety information must not longer be sold. 2. Confirm the receipt of the information with the attached confirmation of receipt form by April 19, 2024. 3. Pass this safety information on to customers of the affected products. 4. Have the receipt of the information confirmed with the attached confirmation of receipt form. Forward the confirmation forms to our attention using the contact information following. For users: 1. If you have a product Urispec 11-Way, #102-2285, LOT 65402, please check if the leucocytes test pad is present before use of the test strips. 2. Pass the information on to other people and departments who may be working with the product Urispec 11-Way, #102-2285, LOT 65402. 3. Confirm the receipt of the information with the attached confirmation of receipt form by April 19, 2024. 4. If your product is affected, please send the product to your local distributor and arrange for replacement or contact MACHEREY-NAGEL directly. Please make sure in your organization that all users of the above products and other persons to be informed become aware of this Urgent Field Safety Notice. Please keep this information at least until all measures have been completed. Contact persons: Contact for sales / marketing: Dr. Sarah Geenen Product Management Rapid Tests Email: sgeenen@mn-net.com Tel.: +49-(0)2421 969 340, Regulatory contact: Dr. Markus Meusel QAM, Manager Regulatory Affairs Email: mmeusel@mn-net.com Tel. +49 (0) 2421 969 210.

K numbers
K991927
Harmonized identifiers (openFDA)
Device name
Method, Enzymatic, Glucose (Urinary, Non-Quantitative)
Medical specialty description
Clinical Chemistry
Regulation number
862.1340
Device class
2
K number
K811816, K843755, K942648, K050801, K931897, K946116, K932363, K864544, K160062, K061846, K022386, K770271, K052525, K861733, K841713, K101673, K160372, K020175, K834387, K063276, K991927, K932368, K843759, K042641, K171083, K932364, K932367, K811817, K131600, K970250, K855239, K091216, K841714, K811818, K180356, K843758, K932369, K023885, K771755, K940469 +41 more
Registration number
3030446844, 3008058135, 3002637618, 2250030, 3038371335, 3011121740, 3006198300, 2432235, 3008517993, 3003422726, 1616487, 3011689956, 3020704087, 3003795116, 9611444, 3002807877, 3006621318, 1810909, 9710094, 9615507, 3003917514, 3005146313, 9610633, 3010324880, 3009238284, 1417592, 9680746, 3010891909, 1181121, 1832216, 3004192065, 1036362, 3012494290, 3012118659, 4580, 3007301207, 1000414342, 3042471840, 3005360469, 3021186226 +17 more
FEI number
3030446844, 3008058135, 3002637618, 2250030, 3038371335, 3006198300, 3002829675, 2432235, 3008517993, 3003422726, 1616487, 3011689956, 3020704087, 3003795116, 3002807877, 3006621318, 1810909, 3003917514, 3005146313, 3010324880, 3009238284, 1417592, 3010891909, 1181121, 1832216, 1000306617, 1000206808, 3004192065, 3010700165, 3002806559, 3012494290, 1036362, 3012118659, 4580, 3007301207, 1000414342, 3042471840, 3021186226, 3003516723, 3005670738 +17 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.