Device recalls

Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order Number G32193; Reference Part N…

Cook Incorporated · Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.

Open, ClassifiedIncorrect or no expiration dateZ-1785-2026
Open FDA's page for this record
Cfres ID
218925
Product res number
Z-1785-2026
Event date initiated
2026-03-05
Event date posted
2026-04-06
Recall status
Open, Classified
Res event number
98495
Product code
BWC
Firm FEI number
1820334
Recalling firm
Cook Incorporated
Address 1
750 N Daniels Way
City
Bloomington
State
IN
Postal code
47404-9120
Additional info contact
Cook Medical Customer Relations Department 812-457-4500
Reason for recall
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause description
Incorrect or no expiration date
Product quantity
476 units
Distribution pattern
Worldwide distribution.
Product description

Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order Number G32193; Reference Part Number C-TCCSB-500-UNI-01, Order Number G47665.

Code info

Reference Part Number C-TCCSB-500-UNI, UDI (01)00827002321939(17)270803(10)NS16189671, Lot Number NS16189671; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191668, Lot Number 16191668; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270807(10)16195472, Lot Number 16195472; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270807(10)16195467, Lot Number 16195467; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191666, Lot Number 16191666; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191677, Lot Number 16191677; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191661, Lot Number 16191661; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270812(10)16200360, Lot Number 16200360; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270731(10)16185929, Lot Number 16185929; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270904(10)16237917, Lot Number 16237917; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191674, Lot Number 16191674; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191671, Lot Number 16191671; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270806(10)16193388, Lot Number 16193388; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270806(10)16193387, Lot Number 16193387; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191667, Lot Number 16191667; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270807(10)16195473, Lot Number 16195473; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191662, Lot Number 16191662; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191689, Lot Number 16191689; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270807(10)16195468, Lot Number 16195468; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270927(10)16271019, Lot Number 16271019.

Action

Cook Medical issued an URGENT: MEDICAL DEVICE RECALL to its consignee on 3/5 2026 via courier or email. The notice explained the issue, potential risk to health, and requested the following: Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.

K numbers
K160200, K010016
Harmonized identifiers (openFDA)
Device name
Needle, Emergency Airway
Medical specialty description
Anesthesiology
Regulation number
868.5090
Device class
2
K number
K153761, K010016, K013241, K790650, K952401, K853345, K160200, K050166, K013252, K771587, K962312, K013916, K130405, K200190
Registration number
3005941719, 3030733800, 3012542015, 1000393132, 2320200, 3012494290, 3030447506, 2025816, 3030446844, 3009039068, 9617604, 3013557562, 8010939, 1820334
FEI number
3002808440, 3030733800, 1000393132, 3012542015, 3012494290, 3030447506, 2025816, 3000203162, 3002558692, 3030446844, 3009039068, 3013557562, 3002807111, 1820334