Device recalls

Brand Name: Contour Product Name: Contour Gold Eyelid Weight Model/Catalog Number: LL3010 Product Description: Contour Gold Eyelid Weight,…

Meddev Corp · Due to mislabeling of products

Open, ClassifiedProcess controlZ-1744-2025
Open FDA's page for this record
Cfres ID
213143
Product res number
Z-1744-2025
Event date initiated
2024-07-12
Event date posted
2025-05-09
Recall status
Open, Classified
Res event number
96551
Product code
NCB
Code info
Model No LL3010; UDI code: 010081314202038217260727107227 Lot Number: 7227
Firm FEI number
2921577
Recalling firm
Meddev Corp
Address 1
730 N Pastoria Ave
City
Sunnyvale
State
CA
Postal code
94085-3522
Additional info contact
Catherine Leong +1-408-7309702
Reason for recall
Due to mislabeling of products
Root cause description
Process control
Product quantity
34 weights
Distribution pattern
U.S. Nationwide distribution in the states of CA, CO, CT, FL, IA, IL, MA, ME, MI, NE, NY, OH, OK, OR, SC, TX, and WV.
Product description

Brand Name: Contour Product Name: Contour Gold Eyelid Weight Model/Catalog Number: LL3010 Product Description: Contour Gold Eyelid Weight, 1.0 gm MedDev Contour and ThinProfile Eyelid Weight Implants are designed for the gravity-assisted treatment of protracted permanent lagophthalmos, usually resulting from facial paralysis.

Action

On 07/12/2024, the firm sent a customer notification letter via first class mail to inform customers that the labels on the Contour Gold Eyelid Implants LL3010/Lot # 7227 potentially do not match the weight; the implants labeled LL3010 (1.0 gm) may contain LL3012 (1.2 gm). Customers are instructed to check their inventory for any/all of the affected products and to contact MedDev Corporation to request for an RGA number in or return affected products and receive replacement products. For questions/concerns contact Quality Compliance Specialist at 1-800-543-2789 or email catherine@meddev-corp.com An updated "URGENT: MEDICAL DEVICE RECALL" Letter was distributed to customers on 04/17/2025. Updated Customer Letter is informing customer that the potential product mix-up may include an additional product and lot number (Model Number LL3012 and Lot Number 7228)

K numbers
K150986
Harmonized identifiers (openFDA)
Device name
Weights, Eyelid, Implantable
Medical specialty description
Ophthalmic
Regulation number
886.5700
Device class
2
K number
K203569
Registration number
9615745, 3002991496, 3004467263
FEI number
3000314405, 3002991496, 3004467263
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.