Device recalls

TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP

TAS Medical Inc · Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties i…

Open, ClassifiedDevice DesignZ-1294-2025
Open FDA's page for this record
Cfres ID
212462
Product res number
Z-1294-2025
Event date initiated
2025-01-27
Event date posted
2025-03-06
Recall status
Open, Classified
Res event number
96274
Product code
GAR
Product description
TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP
Firm FEI number
3029835193
Recalling firm
TAS Medical Inc
Address 1
9100 Conroy Windermere Rd Ste 200 9100 Conroy Windermere Rd
Address 2
Ste 200
City
Windermere
State
FL
Postal code
34786-8431
Root cause description
Device Design
Product quantity
114
Distribution pattern
US Nationwide distribution in the states of LA, OK, TN.
Code info

Model/UDI-DI/Lot(Expiration): T-4000/00197644217267/230403(2024/04/20); T-5000/00860011435921/240801(2025/08/28), 240901(2025/09/25); T-LAP/00860011435914, 00197644927241/240702(2025/07/08), 230906(2024/11/29), 230905(2024/11/29)

Reason for recall

Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.

Action

On 1/27/2025, correction notices were emailed to customers who were asked to do the following: 1) Firm has developed and validated an interim method for conditioning the product in the surgical suite just prior to implantation to rehydrate the zip-tie straps per these instructions. To reverse the effects of dehydration, when preparing for surgery, submerge the TAS zip-tie straps in 40-degreesC (104-degrees F) sterile saline solution for 60 minutes prior to implantation. The orange leader material is not subject to dehydration and therefore is not required to be soaked. 2) Quarantine affected devices and insert the rehydration procedure into each of the remaining product boxes in your inventory at which point it can be released from quarantine. 3) Firm is is in the process of redesigning its packaging to prevent dehydration of the straps and will let you know when the new packaging is available. We anticipate that it may take up to six months to redesign the packaging, validate it, and release it to the field. Until then, the above rehydration procedure will restore the hydration to the straps in the surgical suite thereby preventing breakage that has been seen in the field. 4) Complete and return the acknowledgement and receipt form via email to tkramer@tasmedical.net Questions can be directed to the firm at (716) 997-2896 Monday through Friday between the hours of 8am and 5pm, EST.

K numbers
K220980
Harmonized identifiers (openFDA)
Device name
Suture, Nonabsorbable, Synthetic, Polyamide
Medical specialty description
General, Plastic Surgery
Regulation number
878.5020
Device class
2
K number
K960327, K903177, K003000, K220980, K261484, K922530, K993998, K900531, K030177, K931914, K981322, K930517, K932037, K212888, K904052, K912429, K123034, K990090, K930943, K931613, K943223, K192420, K031531, K926587, K020146, K032160, K042367, K000539, K060471, K943222, K905286, K913102, K024090, K943216, K904931, K001173, K042284, K994176, K151165, K041510 +17 more
PMA number
N83511, N85316, N80838, N85060
Registration number
3030446844, 9616088, 3018094310, 3013298431, 3005953126, 1062741, 3004192065, 3014276972, 9611459, 1111225, 2244478, 3012026312, 2521877, 1061124, 1055236, 3006425876, 3013002341, 3006981798, 3020983460, 3008912461, 1037089, 3008483449, 3010864832, 3004111573, 3011137372, 3004060107, 1066741, 2246552, 9680794, 3015183635, 9612501, 3010131137, 3013548553, 1017294, 3012421607, 3007791573, 1018470, 3030733800, 2523835, 3013403214 +77 more
FEI number
3030446844, 3018094310, 3013298431, 3005953126, 3004192065, 3014276972, 1111225, 3012026312, 2521877, 3001639746, 3006425876, 3013002341, 3006981798, 3020983460, 3008912461, 1037089, 3008483449, 3004111573, 3000206435, 3004060107, 3001236905, 2246552, 3001738911, 3002680121, 3015183635, 3010131137, 3013548553, 1017294, 3002807993, 3012421607, 3007791573, 1018470, 3003737899, 3030733800, 2523835, 3013403214, 3016904853, 3032543753, 3003040625, 3014273254 +77 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.