Device recalls

Brand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software Version: firmware V3.01

LEICA BIOSYSTEMS NUSSLOCH GMBH · Firm became aware of software issues related to most recent firmware update which could result in loss of patient tissue.

Open, ClassifiedUnder Investigation by firmZ-1247-2025
Open FDA's page for this record
Cfres ID
212369
Product res number
Z-1247-2025
Event date initiated
2025-01-30
Event date posted
2025-02-27
Recall status
Open, Classified
Res event number
96220
Product code
IDP
Product description
Brand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software Version: firmware V3.01
Code info
Model No: 1491950C4US UDI-DI: (01)04049188139030 Serial Numbers 12835, 12854
Firm FEI number
3002179321
Recalling firm
LEICA BIOSYSTEMS NUSSLOCH GMBH
Address 1
Heidelberger Str. 17-19 17-19 heidelberger str.
City
Nussloch
Country
Germany
Additional info contact
Margaret Walczak 815-2002717
Reason for recall
Firm became aware of software issues related to most recent firmware update which could result in loss of patient tissue.
Root cause description
Under Investigation by firm
Product quantity
2
Action

On January 30, 2025, the firm notified customers through Email of the recall. Customers were provided with immediate actions to take to mitigate the risk of damaging the tissue sample. Among the firm's recommendations was to have a backup instrument if available. The firm's Service organization will contact customers in the coming weeks to organize the upgrade of affected instruments.

Distribution pattern

Worldwide - US Nationwide distribution in the states of Idaho and Ohio. The countries of France, Australia, Belgium, China, Germany, Spain, UK, Hong Kong, Ireland, Italy, Japan, Netherlands, Norway, Poland, Singapore.

Harmonized identifiers (openFDA)
Device name
Microtome, Cryostat
Medical specialty description
Pathology
Regulation number
864.3010
Device class
1
Registration number
3021013124, 3014680809, 2083544, 1000391282, 8010478, 3005642384, 3043088937, 3043138885, 3027645317, 3003584784, 3010889707, 3008399755, 3006610913, 3043606682, 3004594181, 3015167933, 3010781643, 3038188821, 3002821232, 9615162
FEI number
3021013124, 3014680809, 2083544, 1000391282, 3005642384, 3043088937, 3043138885, 3027645317, 3003584784, 3010889707, 3008399755, 3006610913, 3004063640, 3043606682, 3004594181, 3015167933, 3010781643, 3002179321, 3038188821, 3002821232
K number
K883514
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.