Device recalls

Brivo MR355, NMRI system

GE Healthcare (China) Co., Ltd. · For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display".…

Open, ClassifiedSoftware designZ-1240-2025
Open FDA's page for this record
Cfres ID
212569
Product res number
Z-1240-2025
Event date initiated
2025-01-31
Event date posted
2025-02-28
Recall status
Open, Classified
Res event number
96311
Product code
LNH
Product description
Brivo MR355, NMRI system
Firm FEI number
3008284470
Recalling firm
GE Healthcare (China) Co., Ltd.
Address 1
1F-5F BLD 1, No.1 Tongji South Road, Beijing Economic-Technological Development Area 1f-5f economic-technological development area bld 1 no.1 tongji south road
City
Beijing
Country
China
Additional info contact
GE HealthCare Service 800-4371171
Root cause description
Software design
Product quantity
22
Distribution pattern
Domestic US distribution nationwide. International distribution worldwide.
Code info

Brivo MR355 (No UDI) System IDs 082427010096 082427020046 082427040126 082427070076 082427070077 082427070170 082427100245 082427120151 082427180030 082427210147 082427210148 082427220025 082427220045 082427230060 082427240060 082427310104 0856270058 0856270059 MRR10200 MRR9920 RU1063MR03 RU8902MR01 (Updated 12/29/25 - Additional System ID 83027805544513) (Updated 1/28/26 - Additional System IDs: 82427220047 82427240055 82427070163 82427060068 82427140137 82427010122 82427280018) (Updated 2/19/26: Additional System ID EG1736MR02)

Reason for recall

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

Action

On January 31, 2025, the firm notified customers through Urgent Medical Device Correction letters. Customers were provided with instructions for continued use of their MR systems. All potential users should be made aware of this safety notification and recommended actions. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

K numbers
K123417
Harmonized identifiers (openFDA)
Device name
System, Nuclear Magnetic Resonance Imaging
Medical specialty description
Radiology
Regulation number
892.1000
Device class
2
K number
K924154, K193215, K143345, K945875, K883648, K153740, K202014, K252239, K932344, K091166, K140956, K213693, K935369, K921098, K181370, K974530, K990153, K911975, K250236, K914114, K071925, K052172, K073425, K041695, K223875, K043290, K982932, K910715, K970990, K103411, K920199, K901980, K193282, K211037, K962892, K233673, K900286, K010057, K133848, K113145 +1,070 more
PMA number
P830074, P860010, P840007, P830076, P830053, P860062, P830051, P840063, P850043, P830052, P830081
Registration number
3014635027, 3018094310, 3015223550, 3018423337, 3010326005, 1650907, 3010703925, 3003768277, 3029521953, 3014310698, 3017375271, 3014232555, 3009154005, 3033333119, 3010398867, 2247992, 3023385454, 3006369484, 3002808157, 3010009632, 1054713, 3006015847, 3011657336, 3026047529, 3008284470, 3009196578, 3013343065, 3009755250, 3031503890, 3042177665, 3013679558, 3018004899, 3009714754, 3011015597, 3006239787, 3026391023, 3004977335, 2126677, 3019844073, 3004526608 +110 more
FEI number
3014635027, 3018094310, 3005214683, 3015223550, 3018423337, 3010326005, 3010703925, 3029521953, 3014310698, 3017375271, 3014232555, 3009154005, 1000571342, 3033333119, 3010398867, 3003377038, 3023385454, 3006369484, 3002808157, 3010009632, 3002808382, 3006015847, 3011657336, 3026047529, 3008284470, 3009196578, 3013343065, 1937223, 3009755250, 3031503890, 3042177665, 3013679558, 3018004899, 3009714754, 3011015597, 1000138054, 3006239787, 3026391023, 3004977335, 2126677 +110 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.