Device recalls

DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.

BECKMAN COULTER LABORATORY SYSTEMS (SUZHOU) CO., LTD. · Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assay…

Open, ClassifiedSoftware change controlZ-1233-2025
Open FDA's page for this record
Cfres ID
212289
Product res number
Z-1233-2025
Event date initiated
2025-01-10
Event date posted
2025-02-25
Recall status
Open, Classified
Res event number
96193
Product code
JJE
Product description
DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.
Firm FEI number
3027977508
Recalling firm
BECKMAN COULTER LABORATORY SYSTEMS (SUZHOU) CO., LTD.
Address 1
No. 11 Shengpu Street Zhensheng Road
City
Suzhou
State
Pingjiang
Country
China
Additional info contact
Courtney Walton 952-448-4848
Root cause description
Software change control
Product quantity
56
Code info

Software Versions: V1.4, V1.4.1, V1.4.2, V1.4.3 REF/UDI-DI/Serial Numbers: C63519/14987666545058/2024090170; C63520/14987666545065/2024070116, 2023070009, 2024010038, 2024010035, 2024050072, 2024040056, 2024030049, 2024020046, 2024070111, 2024030051, 2024030052, 2024010033, 2023090015, 2024060103, 2024060104, 2024060101, 2024080145, 2024020045, 2024020042, 2024020043, 2024020047, 2024060096, 2024040059, 2023100019, 2024030053, 2023070010, 2024070129, 2024070130, 2024020041, 2024090183, 2024010036, 2024010037, 2024070132, 2024070131, 2023100020, 2024050064, 2024050070, 2023090016, 2023110022, 2024070110, 2024060075, 2024060106, 2023110023, 2024010024, 2024010025, 2024050066, 2024010029, 2024060107, 2023080012, 2024040060, 2024050074, 2024040062, 2024010031, 2024030055, 2024080144

Reason for recall

Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.

Action

On 1/10/2025, recall notices were mailed to customers who were asked to do the following: 1) Replace and discard any reagent bottles that were on-board the analyzer at the time of any software upgrade. 2) Review Quality Control data and actions taken, for shifts out of allowed laboratory range which could have indicated potential deterioration in reagent stability. 3) Firm recommends sharing the content of this letter with your laboratory and/or Medical Director to determine if a review of previous patient test results should be conducted. 4) If specific software upgrade date is required, contact firm. 5) Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. This issue will not occur with future software updates. If you have any questions regarding this notice, please contact the firm's Customer Support Center: Website: http://www.beckmancoulter.com Hotline: (800) 854-3633

Distribution pattern

US: MA, CA, SC, LA, OK, GA, AL, TX, OR, ID, CO, MN, NY, MS, PA, WA, IL, MI, PR. OUS: Lebanon, Poland, Ghana, Spain, South Africa, Slovakia, Ghana, Croatia, Seychelles, Australia, Republic of Korea

K numbers
K220977
Harmonized identifiers (openFDA)
Device name
Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Medical specialty description
Clinical Chemistry
Regulation number
862.2160
Device class
1
K number
K831649, K840868, K130915, K130276, K882450, K832115, K912844, K922072, K903071, K811919, K881495, K862818, K822465, K860080, K002982, K833448, K843059, K933703, K940810, K842500, K923060, K760642, K810596, K810129, K833833, K921661, K133519, K782014, K840501, K922220, K885226, K862235, K971761, K811194, K901346, K920402, K902336, K950958, K003721, K872494 +168 more
PMA number
P910007
Registration number
9610942, 3005748564, 3005655287, 3010421384, 2026994, 3024082778, 3023162659, 3002642396, 2250051, 3008494306, 9610566, 3029973107, 3000148879, 3003589956, 3004529019, 3006198300, 9614373, 3004973408, 8020888, 9615754, 9680746, 3009518015, 3048199746, 3002809144, 9615837, 3017548799, 3008853996, 3012526002, 3016244760, 3017894300, 1319436, 3005094123, 3039396083, 3003593973, 3027977508, 1319808, 3002895169, 3007111389, 3005333358, 3014717638 +139 more
FEI number
3004445302, 3005748564, 1000305840, 3005655287, 3010421384, 2026994, 3024082778, 3002485711, 3023162659, 3002642396, 2250051, 3008494306, 3029973107, 3000148879, 3003589956, 3004529019, 3002806809, 3006198300, 3004973408, 1000119795, 3009518015, 3048199746, 3002809144, 3002807663, 3017548799, 3008853996, 3012526002, 3016244760, 3017894300, 3002808382, 3005094123, 3002808340, 3002608770, 3039396083, 3003593973, 3027977508, 3002895169, 3007111389, 3005333358, 3014717638 +139 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.