Device recalls

A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 8H LT: Anatomic Lateral Fibula Plate, 8-hole, Left Intended to bridge or otherwise stabilize bone frag…

Tyber Medical · Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibu…

Open, ClassifiedProcess controlZ-1213-2025
Open FDA's page for this record
Cfres ID
212347
Product res number
Z-1213-2025
Event date initiated
2025-01-21
Event date posted
2025-02-21
Recall status
Open, Classified
Res event number
96212
Product code
HRS
Code info
UDI-DI 00196449015566; Lot Numbers 265784, 24043BU01, 261290
Firm FEI number
3012966183
Recalling firm
Tyber Medical
Address 1
83 S Commerce Way Ste 310 83 S Commerce Way
Address 2
Ste 310
City
Bethlehem
State
PA
Postal code
18017-8934
Additional info contact
Zachary Bloshuk 1-866-7610933
Root cause description
Process control
Product quantity
214 units
Distribution pattern
Nationwide
Product description

A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 8H LT: Anatomic Lateral Fibula Plate, 8-hole, Left Intended to bridge or otherwise stabilize bone fragments to facilitate healing. Model/Catalog Number: 770708081

Reason for recall

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Action

Tyber Medical issued Urgent Medical Device Recall letter on 1/24/25 via email to the Distributor. Letter states reason for recall, health risk and action to take: 1. Acknowledge the receipt of this communication via email correspondence to feedback@tybermed.com. 2. Further distribute this communication to all applicable parties and/or customers that possess the impacted inventory. 3. Identify via laser-marked lot number and quarantine the impacted plates under Zimmer Biomet s control that have not been distributed for sale. An RMA will be provided for the return of goods to Tyber Medical for further inspection. 4. Identify via laser-marked lot number and quarantine the impacted plates that have been distributed. An RMA will be provided for the return of goods to Tyber Medical for further inspection. 5. Complete the customer response form located in Appendix 2 and return it to Tyber Medical via email correspondence to feedback@tybermed.com.

K numbers
K232652
Harmonized identifiers (openFDA)
Device name
Plate, Fixation, Bone
Medical specialty description
Orthopedic
Regulation number
888.3030
Device class
2
K number
K910482, K140814, K182201, K061211, K984390, K212608, K233873, K211051, K260323, K011335, K131445, K864924, K193452, K130340, K200538, K202914, K990392, K972913, K111008, K254002, K221465, K234062, K052624, K113648, K090047, K210198, K142943, K143697, K082624, K110908, K192675, K133536, K261892, K071549, K142484, K170780, K213005, K132554, K123203, K103243 +1,292 more
Registration number
3011580264, 1226544, 3023265483, 9614438, 3007051929, 3006801265, 3010047402, 3007993775, 3013280419, 3038503572, 3014527928, 3019923333, 1058084, 3010400367, 9615766, 3010470577, 2032830, 3004187715, 3022777867, 3021336182, 2031009, 3015869121, 3011230048, 3043543260, 3011295718, 3026576110, 3010375065, 3010646402, 2531477, 3036756245, 3034676720, 1218017, 3010039400, 3013540005, 3011127597, 1828288, 1038671, 3005751028, 3027242374, 3009756327 +481 more
FEI number
3011580264, 3023265483, 3007051929, 3006801265, 3010047402, 3007993775, 1000139987, 1000069183, 3013280419, 3038503572, 3007282753, 3014527928, 3019923333, 3002807231, 3002808264, 1000143502, 3003690049, 3010400367, 3010470577, 3004187715, 3022777867, 3008782585, 3021336182, 3015869121, 3011230048, 3043543260, 3011295718, 3026576110, 3010375065, 3010646402, 3002821504, 3036756245, 3034676720, 1218017, 3010039400, 3013540005, 3011127597, 1828288, 1038671, 3005751028 +481 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.