Device recalls

AnchorFast Guard, Oral Endotracheal Tube Fastener, REF 9800,

Hollister Incorporated · Hollister received reports of decreased skin barrier wear time which could can lead to tube migration

Open, ClassifiedProcess change controlZ-1181-2025
Open FDA's page for this record
Cfres ID
212268
Product res number
Z-1181-2025
Event date initiated
2024-12-23
Event date posted
2025-02-19
Recall status
Open, Classified
Res event number
96181
Product code
CBH
Product description
AnchorFast Guard, Oral Endotracheal Tube Fastener, REF 9800,
Code info
UDI/DI 00610075186518, Lot Numbers: 4K162, 4K172, 4K182, 4K192, 4K202, 4K212, 4K222, 4K232, 4K252, 4K262, 4L012, 4L022, 4L032, 4L042, 4L052, 4L132.
Firm FEI number
1480288
Recalling firm
Hollister Incorporated
Address 1
2000 Hollister Dr
City
Libertyville
State
IL
Postal code
60048-3781
Additional info contact
Christine Ponka 224-358-5642
Reason for recall
Hollister received reports of decreased skin barrier wear time which could can lead to tube migration
Root cause description
Process change control
Product quantity
4,955 box / 59,460 eaches
Distribution pattern
Worldwide distribution - US Nationwide and the country of Canada.
Action

Hollister issued an URGENT - MEDICAL DEVICE RECALL notice to its consignees on 12/23/2024 via FedEx and email. The notice explained the reason for the recall and potential risk and instructed customers to quarantine affected inventory, discard the product per their local procedures for product destruction, and Please complete and return the response form attached as soon as possible, even if you do not have affected product in your possession and send to ComplaintCommunication@Hollister.com. For product credit requests: If you are a direct customer of Hollister, contact Hollister Incorporated Customer Service for product credit at: 1-800-323-4060 (prompt #1). If you purchase product from a distributor, please contact your distributor for product credit. Hollister has posted this recall on their website with the link to the page.

Harmonized identifiers (openFDA)
Device name
Device, Fixation, Tracheal Tube
Medical specialty description
Anesthesiology
Regulation number
868.5770
Device class
1
K number
K822794, K903539, K882079, K864526, K862084, K811185, K922657, K882694, K860845, K882891, K843790, K872832, K830706, K900912, K936094, K802601, K942529, K810737, K904974, K942395, K781695, K872277, K940448, K941384, K900787, K950318, K934288, K933781, K912884, K851874, K871560, K935171, K942016, K902604, K902065, K905231, K900236, K964399, K921698, K950930 +7 more
Registration number
1214422, 3003965134, 3015416207, 3013298431, 3021234365, 9611024, 3015058854, 3006113262, 8010939, 8040412, 1061124, 9616567, 1314417, 3013550220, 3006755899, 8021995, 3016112560, 3011600579, 3011137372, 3006157842, 1526863, 3003902312, 3013412170, 3006563173, 3004727093, 3010834934, 3034669683, 3013735622, 3007483335, 1647162, 3026036971, 3016171961, 1921846, 3008311916, 8010485, 3014146451, 3010131137, 2025816, 1417592, 3045058049 +107 more
FEI number
3003361161, 1214422, 3003965134, 3015416207, 3013298431, 3021234365, 3015058854, 3006113262, 3002807170, 1000306115, 1314417, 3013550220, 3005304066, 3002806816, 3006755899, 3016112560, 3011600579, 3006157842, 3001236905, 1526863, 3003902312, 3013412170, 1000125536, 3006563173, 3010834934, 3004727093, 3034669683, 3013735622, 3007483335, 3026036971, 3016171961, 1921846, 3008311916, 3014146451, 3010131137, 2025816, 1417592, 3045058049, 1220850, 3011248163 +107 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.