Device recalls

TERUMO HydroPearl Compressible Microspheres for Embolisation REF 8HP2S600

MICROVENTION INC. · Due to, during the manufacturing process, the prescribed manufacturing and quality processes were not followed and the product was inadvertently distributed.

Open, ClassifiedDevice DesignZ-0868-2024
Open FDA's page for this record
Cfres ID
205148
Product res number
Z-0868-2024
Event date initiated
2023-12-21
Event date posted
2024-01-31
Recall status
Open, Classified
Res event number
93692
Product code
NOY
Product description
TERUMO HydroPearl Compressible Microspheres for Embolisation REF 8HP2S600
Code info
Model/Catalog Number: HP2S0600 UDI Code: (01)00816777026118(11)230710(17)260630(10)0000342047 Lot Number: 0000342047
Firm FEI number
3013556777
Recalling firm
MICROVENTION INC.
Address 1
35 Enterprise
City
Aliso Viejo
State
CA
Postal code
92656
Additional info contact
Terrence Callahan 949-899-7194
Reason for recall
Due to, during the manufacturing process, the prescribed manufacturing and quality processes were not followed and the product was inadvertently distributed.
Root cause description
Device Design
Product quantity
22 units
Distribution pattern
U.S. Nationwide distribution in the states of AZ, IL, LA, NC, TN, VA, and WI.
Action

On 12/21/2023, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter via FedEx Overnight Certified Letter to customers informing them that MicroVention is recalling one lot (#00003420470 of HYDRPEARL Microsphere Device (HP2S0600) due to the firm no following the prescribed manufacturing and quality processes and the affected product was inadvertently released for distribution. Specifically, the in-processing ageing step was mistakenly extended an additional five (5) days. Customers are instructed to: 1. Immediately stop using and quarantine the impacted HYDROPEARL device. 2. Inform all individuals with their organization of the recall and forward to any organizations that they may have received the affected products. 3. Complete the CUSTOMER ACKNOWLEDGEMENT FORM to gina.digioia@terumomedical.com 4. If their institution has any affected inventory, contact Customer Care at 800-888-3786 or email tmccustomer.admin@terumomedical.com to request a recall return and received RGA details to return the device(s). For questions or assistance, contact Director, Post-Market Surveillance at terrence.callahan@microvention.com

K numbers
K150870, K192684
Harmonized identifiers (openFDA)
Device name
Agents, Embolic, For Treatment Of Benign Prostatic Hyperplasia
Medical specialty description
Gastroenterology, Urology
Regulation number
876.5550
Device class
2
K number
K203276, DEN160040, K180102, K192684
Registration number
1225687, 2134265, 9616684, 3005111382, 3013556777, 3015309643, 9615728, 3033589330, 2124215, 3012536737, 3008262715
FEI number
3002808761, 3005111382, 3013556777, 3002095335, 3001451463, 3015309643, 3033589330, 2124215, 3012536737, 3008262715, 3002913962
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.