Device recalls

Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of moveme…

Open, ClassifiedUnder Investigation by firmZ-0861-2024
Open FDA's page for this record
Cfres ID
205331
Product res number
Z-0861-2024
Event date initiated
2023-12-08
Event date posted
2024-01-30
Recall status
Open, Classified
Res event number
93734
Product code
OWB
Product description
Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
Firm FEI number
3002807880
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Address 1
Veenpluis 4-6 4-6 veenpluis
City
Best
Country
Netherlands
Root cause description
Under Investigation by firm
Product quantity
108 systems (14 US, 1 Canada, 93 ROW)
Distribution pattern
Worldwide - US Nationwide distribution in the states of CO, FL, IL, MS, NV and TX. The country of Canada.
Code info

(1) System Code: 722079; (1) UDI/DI: 00884838085268; (1) Serial Number: 1005, 866, 1361, 952, 1241, 1169, 756, 703351, 1135, 1122, 824, 748, 630, 638, 703385, 812, 703346, 1303, 1313, 893, 768, 929, 703638, 1357, 703810, 934, 1202, 703461, 703706, 1172, 703959, 703845, 703705; (2) System Code: 722224; (2) UDI/DI: 00884838099258; (2) Serial Number: 539, 395, 278, 134, 730, 252, 370, 36, 572, 645, 272, 495, 781, 62, 1388, 352, 1655, 690, 555, 895, 2165, 861, 60, 20, 648, 2231, 1778, 1431, 778, 477, 917, 1718, 388, 2042, 150, 764, 2036, 1421, 157, 879, 212, 144, 1463, 347, 575, 2187, 355, 2576, 1332, 424, 387, 863, 211, 474, 712, 549, 2309, 799, 2283, 639, 1700, 2071, 537, 1589, 2208, 2057, 771, 772, 1216, 1667;

Reason for recall

Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.

Action

The recalling firm issued an URGENT: MEDICAL DEVICE CORRECTION notification dated 12/08/2023 by certified mail to consignees. Consignees with affected devices are asked to utilize the provided workaround instructions. Additionally, consignees are asked to make all users aware of this recall notification and keep a copy of this notification with system documentation. Getinge Field Service Engineers will be dispatched to the field to install software updates to affected devices. Consignees with any questions can contact the recalling firm by phone at 1-800-722-9377. Consignees are asked to return the provided response form to the recalling firm at ITG_Recalls@philips.com.

K numbers
K200917
Harmonized identifiers (openFDA)
Device name
Interventional Fluoroscopic X-Ray System
Medical specialty description
Radiology
Regulation number
892.1650
Device class
2
K number
K243735, K170130, K253670, K221344, K220342, K173631, K180473, K243884, K211191, K200917, K050489, K211780, K212442, K231700, K101868, K181550, K173314, K251523, K163715, K123508, K181804, K240465, K032046, K231669, K130638, K172646, K062647, K203281, K052202, K181834, K002198, K212383, K033244, K230787, K160147, K961374, K123529, K181415, K222625, K233230 +285 more
Registration number
3021059265, 3016438694, 3014391847, 2027467, 3038614464, 3010398867, 9681146, 3004785967, 3015223550, 3009196578, 3008632256, 3003202425, 3012761544, 1723170, 3001722928, 2243057, 3003768251, 3021561977, 3010703925, 3036666444, 3021027284, 3013184485, 3022378058, 3008284470, 3011713200, 3027373189, 3036255956, 3018094310, 8030233, 3006544299, 3016248538, 1225677, 3009973505, 1037955, 2133772, 3008363989, 3025394757, 2939520, 3030412764, 3014343100 +86 more
FEI number
3021059265, 3016438694, 1000114111, 3014391847, 2027467, 3038614464, 3002808139, 3010398867, 3004785967, 3015223550, 3009196578, 3008632256, 3003202425, 3012761544, 3001722928, 2243057, 3002980729, 3003768251, 3021561977, 3010703925, 3036666444, 3021027284, 3013184485, 3022378058, 3008284470, 3011713200, 3027373189, 3036255956, 3018094310, 3006544299, 3016248538, 3009973505, 1037955, 1000132370, 3008363989, 3025394757, 2939520, 3030412764, 3014343100, 1000188474 +86 more