Generic Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System Model/Catalog Number: 9880CM29A Softwar…
Edwards Lifesciences, LLC · Due to a manufacturing issue which may result in cardiovascular balloon rupturing.
Open FDA's page for this record- Cfres ID
- 223232
- Product res number
- Z-0287-2027
- Event date initiated
- 2026-10-05
- Event date posted
- 2026-11-02
- Recall status
- Open, Classified
- Res event number
- 100067
- Product code
- NPU
- Code info
- Lot Code: Model No.: 9880CM29A; UDI-DI: 00690103219777; Recalled Lot Number(s): 67718561, 67804014, 67887043, 67887044.
- Firm FEI number
- 2015691
- Recalling firm
- Edwards Lifesciences, LLC
- Address 1
- 1 Edwards Way
- City
- Irvine
- State
- CA
- Postal code
- 92614-5688
- Additional info contact
- Brian Hudson +1-949-2503569
- Reason for recall
- Due to a manufacturing issue which may result in cardiovascular balloon rupturing.
- Root cause description
- Process change control
- Product quantity
- 164
- Distribution pattern
- U.S.: AZ, CA, CO, FL, GA, IL, KS, LA, MA, MI, MN, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, UT, VA and WA. O.U.S.: N/A
- Product description
Generic Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System Model/Catalog Number: 9880CM29A Software Version: Not applicable. Product Description: Prosthesis, mitral valve, percutaneously delivered Component: No
- Action
On 10/05/2026, an "URGENT: MEDICAL DEVICE RECALL (PRODUCT REMOVAL)" Letter was sent via various delivery methods (e.g., electronic delivery (e-mail), FedEx Priority Overnight mail, or on-site visit) to customers informing them that, due to complaints certain manufacturing lots of the SAPIEN M3 Commander M delivery systems potentially contain a manufacturing defect, which may result in balloon burst during use. Customers are instructed to: " Review this notice carefully and share it with the appropriate clinical staff at their site. " Immediately examine their inventory and quarantine units subject to this recall. Do not use these affected units. Units with other lot numbers are not impacted by this recall and can continue to be used. " Complete the attached Customer Acknowledgment Form and return it to Edwards TMTT Representative or via email to US_FCA@edwards.com " Contact Customer Service at 1-800-424-3278 or email: CustomerServ_TMTT@edwards.com to obtain a Returned Goods Authorization (RGA) number and return any impacted product to Edwards. Edwards asks that customers complete the last two actions listed above within 15 days of receipt of this notification. Customers' assistance is crucial to ensuring this recall is executed thoroughly and promptly. This notice has been communicated by Edwards to the appropriate Regulatory Authorities. For questions and further assistance, contact an Edwards representative who can assist with product return.
- PMA numbers
- P250019
- Harmonized identifiers (openFDA)
- Device name
- Prosthesis, Mitral Valve, Percutaneously Delivered
- Medical specialty description
- Unknown
- Device class
- 3
- PMA number
- P250019, P140031, P240042
- Registration number
- 3010432890, 3007063218, 1928237, 2015691, 3008500478, 3012179728, 2135147, 3005941719, 3015287573, 3009420598, 3002807314, 3014918977, 3017298588, 3005012805, 3010421104, 3007113487, 2011171, 1721676
- FEI number
- 1713910, 3010432890, 3005334138, 3007063218, 1928237, 3012179728, 3002807111, 3015287573, 3005012805, 3017298588, 3009420598, 3002807314, 3014918977, 3010421104, 2015691, 3007113487, 2011171, 1721676