Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/130; Material Number: M001195210
Boston Scientific Cork, Ltd. · The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcathe…
Open FDA's page for this record- Cfres ID
- 222901
- Product res number
- Z-0224-2027
- Event date initiated
- 2026-09-16
- Event date posted
- 2026-10-22
- Recall status
- Open, Classified
- Res event number
- 99906
- Product code
- KRA
- Product description
- Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/130; Material Number: M001195210
- Firm FEI number
- 3002913962
- Recalling firm
- Boston Scientific Cork, Ltd.
- Address 1
- Boston Scientific Limited, Business And Technology Park, Model Farm Road Boston Scientific Limited Business And Technology Park Model Farm Road
- City
- Cork
- Country
- Ireland
- Additional info contact
- Clara Johnson 763-494-1133
- Root cause description
- Process control
- Product quantity
- 2547
- Code info
Material Number: M001195210; UDI-DI: 08714729839323; Batch Numbers: 37914435, 37914436, 37956674, 37981343, 37981344, 38022632, 38022633, 38022634, 38149673, 38181211, 38224528, 38259475, 38259476, 38290713, 38290714, 38290715, 38290716, 38290717, 38399232, 38446223, 38479515, 38506196, 38530446, 38530447, 38629332, 38629333, 38647617, 38647618, 38649501, 38649508, 38687327, 38708440, 38708441, 38745184, 38769513, 38769514, 38934430, 39189795, 39189798, 40466471, 40466472, 40482683, 40482684, 40482685, 40487838, 40487839, 40487900, 40487954, 40487955, 40487956, 40487957, 40501048, 40501049, 40514122, 40514123, 40514222, 40514223, 40533342, 40533344, 40536165, 40536166;
- Reason for recall
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
- Action
On September 16, 2026, Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. Actions to be taken: 1. Immediately stop further use or distribution and segregate affected product (see Attachment 1 for identification of the affected devices). 2. Forward this notice to relevant personnel and any facilities to which the product was transferred. 3. Continue to follow the standard of practice when monitoring patients treated with the product. 4. Complete and return the enclosed Reply Verification Tracking Form per the included instructions. 5. Return product to Boston Scientific in accordance with the enclosed instructions.
- Distribution pattern
Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom.
- K numbers
- K163701
- Harmonized identifiers (openFDA)
- Device name
- Catheter, Continuous Flush
- Medical specialty description
- Cardiovascular
- Regulation number
- 870.1210
- Device class
- 2
- K number
- K030688, K896819, K161962, K072724, K935349, K081147, K071468, K093750, K132947, K111490, K974473, K201907, K954949, K945593, K123477, K812173, K962129, K032494, K170931, K994155, K961668, K161967, K040139, K182071, K213442, K801829, K210339, K972518, K052841, K905447, K770432, K034036, K020680, K211110, K191305, K043538, K070279, K080737, K964154, K050630 +137 more
- Registration number
- 3040099838, 3002806337, 3031823122, 3017823876, 3005941719, 3014498720, 2032112, 3008824097, 3009380063, 1000121056, 3010334444, 3004415014, 1220948, 2134812, 3045042524, 3006950086, 3003120897, 3014590708, 3011137372, 3012536737, 1718873, 1721686, 3000126629, 1319639, 1644312, 8010026, 2183870, 3005160949, 3009756153, 1721676, 1721504, 2183744, 1724474, 3017208642, 1928237, 1319211, 3010909876, 2024311, 9615817, 3014758755 +109 more
- FEI number
- 3040099838, 3002806337, 3031823122, 3017823876, 3014498720, 3008824097, 3009380063, 1000121056, 3010334444, 3004415014, 1220948, 3045042524, 3006950086, 3002534070, 3003120897, 3014590708, 3012536737, 1718873, 1721686, 3008133910, 3000126629, 3002806565, 2183870, 3005160949, 3009756153, 1721676, 2828, 3005747797, 1721504, 1000220862, 2183744, 3002806476, 3017208642, 1928237, 1319211, 3010909876, 2024311, 3014758755, 3013556777, 3015151147 +109 more