Device recalls

Ion Torrent Genexus Dx with software, Catalog: A40003539, which supports the Ion Torrent Genexus Dx Integrated Sequencer

Life Technologies Holdings Pte Ltd · DNA or RNA sequencing system software has vulnerabilities.

Open, ClassifiedSoftware designZ-0211-2027
Open FDA's page for this record
Cfres ID
222804
Product res number
Z-0211-2027
Event date initiated
2026-08-14
Event date posted
2026-10-22
Recall status
Open, Classified
Res event number
99885
Product code
PFF
Product description
Ion Torrent Genexus Dx with software, Catalog: A40003539, which supports the Ion Torrent Genexus Dx Integrated Sequencer
Firm FEI number
3003673482
Recalling firm
Life Technologies Holdings Pte Ltd
Address 1
Blk 33 Marsiling Ind Estate Road 3 #07-06
City
Singapore
Country
Singapore
Additional info contact
6362-9300
Reason for recall
DNA or RNA sequencing system software has vulnerabilities.
Root cause description
Software design
Product quantity
16
Distribution pattern
US Nationwide distribution in the states of PA, NC, CO, CA, LA, KY, NY, AZ, MA, LA, OK.
Code info

UDI-DI: 10190302018883, Software: 6.8.2.1 and earlier. Sequencer UDI-DI: 10190302017794. Sequencer Serial Numbers: 2619124040913, 2619726010247, 2619124060941, 2619124060942, 2619725090229, 2619725100238, 2619122070666, 2619725080220, 2619725090226, 2619726010251, 2619726020265, 2619725090225, 2619725100235, 2619725050205, 2619726010249, 2619122070658

Action

On 8/14/2026, recall notices were emailed to customers who were informed of the following: Once the software becomes available, your field support specialist or your local Service and Support representative will reach out to schedule a software upgrade. Firm recommends that customers back up their data before any software upgrades. Please refer to "Manage backup settings and restore runs (administrator)" section of the user guide. Complete and return the CUSTOMER RESPONSE Acknowledgement and Receipt Form via email to genexusdx.notifications@thermofisher.com; and also contact the firm at this email address is you have any technical questions or concerns.

Harmonized identifiers (openFDA)
Device name
High Throughput Dna Sequence Analyzer
Medical specialty description
Immunology
Regulation number
862.2265
Device class
2
K number
DEN130011, K170299
Registration number
3015529000, 3009875001, 3006296346, 3007287152, 3003673482, 3007102730, 3011104381, 3003335080, 3003218906, 3014150341, 3009207631
FEI number
3015529000, 3009875001, 3007287152, 3003673482, 3007102730, 3011104381, 3003335080, 3003218906, 3014150341, 3005095756, 3009207631
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.