Device recalls

Techno-Path Multichem AE Unassayed Level 2 REF: AE622X

TECHNO-PATH MANUFACTURING LTD. · A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. A…

Open, ClassifiedUnder Investigation by firmZ-0206-2027
Open FDA's page for this record
Cfres ID
222305
Product res number
Z-0206-2027
Event date initiated
2026-08-14
Event date posted
2026-10-22
Recall status
Open, Classified
Res event number
99724
Product code
OHQ
Product description
Techno-Path Multichem AE Unassayed Level 2 REF: AE622X
Code info
UDI:(01)05391523442817/Lot # 65705262
Firm FEI number
3008156187
Recalling firm
TECHNO-PATH MANUFACTURING LTD.
Address 1
Fort Henry Business Park
City
Ballina Tipperary
Country
Ireland
Additional info contact
Bernd Hass +353-61525700
Root cause description
Under Investigation by firm
Product quantity
123 units
Distribution pattern
US Nationwide distribution in the state of Massachusetts.
Reason for recall

A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.

Action

On August 14, 2026, Techno Path Manufacturing Ltd has issued a medical device correction to consignees via email. Techno Path Manufacturing Ltd informed consignees to please verify their inventory against the affected product codes and lot numbers. Techno Path Manufacturing Ltd. has informed the consignees to take the following actions: 1. Immediately place all affected inventory on hold and prevent further distribution. 2. Quarantine all affected product currently under your control. 3. Confirm that there has been no distribution of the impacted lots to customers. 4. Confirm receipt of this notification and provide inventory and distribution details to Technopath Manufacturing Ltd. within 3 business days. 5. Arrange return or destruction of affected inventory in accordance with instructions provided by Technopath Manufacturing Ltd.

Harmonized identifiers (openFDA)
Device name
Multi-Analyte Controls Unassayed
Medical specialty description
Clinical Chemistry
Regulation number
862.1660
Device class
1
Registration number
3003610890, 1319681, 3004175381, 3014607865, 1050155, 3010934853, 9681518, 2250051, 3010939897, 3003825930, 8043909, 3012963943, 3008156187, 1524213, 3013887826, 1627497, 8020790, 3012918440, 3021090658, 3003855297, 3005959679, 3004530258, 3007088335, 3027518847, 3003356736, 3008386529, 1643621, 3004493545, 3004737529, 3013508066, 2432495, 9610566, 3027062375, 2085064, 2023365, 3009672007, 3002642396, 3032324401, 3050015, 3000202849 +18 more
FEI number
3003610890, 3004175381, 3014607865, 1050155, 3010934853, 1000122536, 2250051, 3010939897, 3003825930, 1220394, 3012963943, 1627497, 3008156187, 1524213, 3013887826, 3002807876, 3012918440, 3021090658, 3003855297, 3005959679, 3004530258, 1000136573, 1000135102, 3027518847, 3003356736, 3008386529, 1643621, 3004493545, 3004737529, 3013508066, 2432495, 1318994, 3027062375, 2023365, 3009672007, 3002642396, 3003407015, 3032324401, 3050015, 3000202849 +18 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.