Device recalls

CareLink Clinic, REF: MMT-7350

Medtronic MiniMed, Inc. · CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values e…

Open, ClassifiedSoftware designZ-0203-2027
Open FDA's page for this record
Cfres ID
222434
Product res number
Z-0203-2027
Event date initiated
2026-09-11
Event date posted
2026-10-21
Recall status
Open, Classified
Res event number
99765
Product code
PHV
Product description
CareLink Clinic, REF: MMT-7350
Code info
UDI-DI: 0763000B00008748K. Software Versions: 4.4A Software Available: US: Between June 4th, 2026, and June 8th, 2026.
Firm FEI number
3003166194
Recalling firm
Medtronic MiniMed, Inc.
Address 1
18000 Devonshire St
City
Northridge
State
CA
Postal code
91325-1219
Additional info contact
Ashley Patterson 800-646-4633
Root cause description
Software design
Product quantity
1
Distribution pattern
US Nationwide distribution in the state of KS.
Reason for recall

CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).

Action

On 9/11/2026, correction notices were mailed to the health care professional, and customer asking them to do the following: The Customer was asked to do the following: Immediately contact your Healthcare Provider (HCP) to confirm your insulin settings values. Complete and return the confirmation form by visiting https://info.medtronicdiabetes.com/FA1587 The health care professional was asked to do the following: If you provided diabetes treatment recommendations to patients based on custom Total Daily Dose(TDD) or long-acting insulin values entered into the CareLin Clinic Settings Calculator between: June 04, 2026, to June 08, 2026, please revisit those recommendations utilizing patients' individual CareLink reports. Continue to follow up with those patients pursuant to your standard of care. If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 1-800-646-4633, option 1.

Harmonized identifiers (openFDA)
Device name
Continuous Glucose Monitor Retrospective Data Analysis Software
Medical specialty description
Clinical Chemistry
Regulation number
862.2120
Device class
1
K number
DEN140016
Registration number
3009862700, 3021611414, 3008770957, 3014642500, 2032227, 2954323, 3012128418, 3028170577, 3019004087, 3013922904, 3004753838, 3024580094, 3009137498, 3007981285, 3014585508
FEI number
3009862700, 3021611414, 3008770957, 3014642500, 3012128418, 3028170577, 3019004087, 3003166194, 3002803444, 3013922904, 3004753838, 3024580094, 3009137498, 3007981285, 3014585508