CareLink Clinic, REF: MMT-7350
Medtronic MiniMed, Inc. · CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values e…
Open FDA's page for this record- Cfres ID
- 222434
- Product res number
- Z-0203-2027
- Event date initiated
- 2026-09-11
- Event date posted
- 2026-10-21
- Recall status
- Open, Classified
- Res event number
- 99765
- Product code
- PHV
- Product description
- CareLink Clinic, REF: MMT-7350
- Code info
- UDI-DI: 0763000B00008748K. Software Versions: 4.4A Software Available: US: Between June 4th, 2026, and June 8th, 2026.
- Firm FEI number
- 3003166194
- Recalling firm
- Medtronic MiniMed, Inc.
- Address 1
- 18000 Devonshire St
- City
- Northridge
- State
- CA
- Postal code
- 91325-1219
- Additional info contact
- Ashley Patterson 800-646-4633
- Root cause description
- Software design
- Product quantity
- 1
- Distribution pattern
- US Nationwide distribution in the state of KS.
- Reason for recall
CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).
- Action
On 9/11/2026, correction notices were mailed to the health care professional, and customer asking them to do the following: The Customer was asked to do the following: Immediately contact your Healthcare Provider (HCP) to confirm your insulin settings values. Complete and return the confirmation form by visiting https://info.medtronicdiabetes.com/FA1587 The health care professional was asked to do the following: If you provided diabetes treatment recommendations to patients based on custom Total Daily Dose(TDD) or long-acting insulin values entered into the CareLin Clinic Settings Calculator between: June 04, 2026, to June 08, 2026, please revisit those recommendations utilizing patients' individual CareLink reports. Continue to follow up with those patients pursuant to your standard of care. If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 1-800-646-4633, option 1.
- Harmonized identifiers (openFDA)
- Device name
- Continuous Glucose Monitor Retrospective Data Analysis Software
- Medical specialty description
- Clinical Chemistry
- Regulation number
- 862.2120
- Device class
- 1
- K number
- DEN140016
- Registration number
- 3009862700, 3021611414, 3008770957, 3014642500, 2032227, 2954323, 3012128418, 3028170577, 3019004087, 3013922904, 3004753838, 3024580094, 3009137498, 3007981285, 3014585508
- FEI number
- 3009862700, 3021611414, 3008770957, 3014642500, 3012128418, 3028170577, 3019004087, 3003166194, 3002803444, 3013922904, 3004753838, 3024580094, 3009137498, 3007981285, 3014585508