ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846…
Beckman Coulter, Inc. · Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break int…
Open FDA's page for this record- Cfres ID
- 222799
- Product res number
- Z-0202-2027
- Event date initiated
- 2026-09-21
- Event date posted
- 2026-10-21
- Recall status
- Open, Classified
- Res event number
- 99882
- Product code
- JQC
- Firm FEI number
- 3007448124
- Recalling firm
- Beckman Coulter, Inc.
- Address 1
- 5355 W 76th St
- City
- Indianapolis
- State
- IN
- Postal code
- 46268-4166
- Additional info contact
- Kristin Godfredsen 760-419-7078
- Root cause description
- Nonconforming Material/Component
- Product quantity
- 336 units
- Distribution pattern
- US Nationwide distribution.
- Product description
ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
- Code info
Serial Numbers: 24U15250 to 26U15766; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
- Reason for recall
Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
- Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 9/9/26 was sent to customers. Return the completed customer response form to Beckman Coulter. Upon receipt of your form, we will coordinate a replacement of your rotor(s). If you have another compatible rotor available that is not included in the Serial Numbers listed above, discontinue use of the affected rotor(s) and use the alternative rotor until your replacement rotor is received. A description of the compatible rotors is available in the instrument instructions for use. If no alternative compatible rotor is available and continued use is necessary, you may continue to use your current rotor until you receive a replacement rotor; however, continued use carries a possibility of rotor breakage which may result in significant damage to materials (i.e., samples) being run in the centrifuge and the centrifuge instrument. Should you experience rotor breakage, please contact your service representative. Use personal protective equipment, and standard laboratory operating procedures for the handling of biohazardous materials when opening the centrifuge to prevent exposure to damaged materials (i.e., samples). Once your completed response form is received, a Beckman Coulter representative will coordinate a free-of-charge replacement order for you. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via electronic system. " Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact Customer Support Center. "From our website: http://www.beckman.com "By
- Harmonized identifiers (openFDA)
- Device name
- Centrifuges (Micro, Ultra, Refrigerated) For Clinical Use
- Medical specialty description
- Clinical Chemistry
- Regulation number
- 862.2050
- Device class
- 1
- K number
- K840768, K030340, K831948, K822715, K861146, K843087, K822422, K802210, K821756, K021902, K052925, K760467, K024080, K042473, K782124, K103340, K183205, K831947, K843088, K030555, K812505, K070666, K991430, K881254, K843434, K850485, K864974, K994148, K761013, K071934, K831942, K800381, K944738, K081345, K860747, K860172, K813524, K810833
- Registration number
- 3012481754, 3011544811, 3015651306, 9616088, 9680411, 2249760, 3006697110, 9611274, 3010704506, 3009711478, 3021013124, 3002579136, 3015086789, 8030607, 3021336182, 3007712511, 3027339877, 3012101664, 3005743849, 8010707, 1066491, 3009034535, 3010394461, 3010400367, 3009746061, 3010381606, 3010331645, 3015505238, 3009756153, 8010672, 3007123908, 1721676, 3030040496, 2032521, 9617426, 2242368, 3028019212, 3004488394, 3011765035, 3007895168 +126 more
- FEI number
- 1000268645, 3011544811, 3015651306, 5833, 3006697110, 3010704506, 3009711478, 3021013124, 3002579136, 3015086789, 3002902735, 3021336182, 3002808075, 3002808043, 3000215273, 3007712511, 3002935209, 3027339877, 3012101664, 3005743849, 3000200590, 3009034535, 3010394461, 3010400367, 3009746061, 3010381606, 3002807091, 3010331645, 3015505238, 3009756153, 3007123908, 1721676, 3030040496, 2242368, 3028019212, 3000203293, 3004488394, 3011765035, 3007895168, 3004063640 +126 more