ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846…

Beckman Coulter, Inc. · Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifuga…

Open, ClassifiedNonconforming Material/ComponentZ-0201-2027
Open FDA's page for this record
Cfres ID
222797
Product res number
Z-0201-2027
Event date initiated
2026-09-21
Event date posted
2026-10-21
Recall status
Open, Classified
Res event number
99882
Product code
JQC
Firm FEI number
3007448124
Recalling firm
Beckman Coulter, Inc.
Address 1
5355 W 76th St
City
Indianapolis
State
IN
Postal code
46268-4166
Additional info contact
Kristin Godfredsen 760-419-7078
Root cause description
Nonconforming Material/Component
Product quantity
92 units
Distribution pattern
US Nationwide distribution.
Product description

ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)

Code info

Serial Numbers: 24U15260 to 26U15764; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860

Reason for recall

Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.

Action

An URGENT MEDICAL DEVICE RECALL notification letter dated 9/9/26 was sent to customers. Return the completed customer response form to Beckman Coulter. Upon receipt of your form, we will coordinate a replacement of your rotor(s). If you have another compatible rotor available that is not included in the Serial Numbers listed above, discontinue use of the affected rotor(s) and use the alternative rotor until your replacement rotor is received. A description of the compatible rotors is available in the instrument instructions for use. If no alternative compatible rotor is available and continued use is necessary, you may continue to use your current rotor until you receive a replacement rotor; however, continued use carries a possibility of rotor breakage which may result in significant damage to materials (i.e., samples) being run in the centrifuge and the centrifuge instrument. Should you experience rotor breakage, please contact your service representative. Use personal protective equipment, and standard laboratory operating procedures for the handling of biohazardous materials when opening the centrifuge to prevent exposure to damaged materials (i.e., samples). Once your completed response form is received, a Beckman Coulter representative will coordinate a free-of-charge replacement order for you. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via electronic system. " Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact Customer Support Center. "From our website: http://www.beckman.com "By

Harmonized identifiers (openFDA)
Device name
Centrifuges (Micro, Ultra, Refrigerated) For Clinical Use
Medical specialty description
Clinical Chemistry
Regulation number
862.2050
Device class
1
K number
K802210, K021902, K030340, K782124, K183205, K944738, K081345, K860172, K831948, K994148, K024080, K103340, K850485, K991430, K822715, K042473, K831942, K052925, K840768, K831947, K860747, K843434, K864974, K843087, K810833, K070666, K822422, K881254, K030555, K800381, K812505, K071934, K813524, K821756, K843088, K760467, K761013, K861146
Registration number
2249760, 3005743849, 3019387954, 3027339877, 2025606, 3004111573, 3009756153, 8010672, 1616487, 3011260354, 3014325803, 3002821232, 3030516433, 3009423112, 3012481754, 8010707, 3006788678, 3003559191, 3005271071, 3006404071, 3012494290, 3013510285, 9611274, 1836161, 3011966385, 3003799895, 1423662, 3045058049, 3019269298, 1450662, 3010386080, 3028019212, 3004417597, 3030178466, 3009711478, 3015086789, 3003965134, 3014657561, 3010394461, 1066491 +126 more
FEI number
3005743849, 3019387954, 1000268645, 3027339877, 2025606, 3004111573, 3009756153, 3003429569, 1616487, 3011260354, 3014325803, 3002821232, 3030516433, 3009423112, 3006788678, 3003559191, 3005271071, 3006404071, 3012494290, 3013510285, 3011966385, 3003799895, 3045058049, 3019269298, 1450662, 3010386080, 3028019212, 1000512168, 3004417597, 3030178466, 3009711478, 3015086789, 3003965134, 3014657561, 3010394461, 3030040496, 3013767421, 3009746061, 3002935209, 3008205331 +126 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.