Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood C…
Greiner Bio-One North America, Inc. · Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) a…
Open FDA's page for this record- Cfres ID
- 221731
- Product res number
- Z-0200-2027
- Event date initiated
- 2026-07-29
- Event date posted
- 2026-10-21
- Recall status
- Open, Classified
- Res event number
- 99518
- Product code
- JKA
- Code info
- Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12
- Firm FEI number
- 3001451379
- Recalling firm
- Greiner Bio-One North America, Inc.
- Address 1
- 4238 Capital Dr
- City
- Monroe
- State
- NC
- Postal code
- 28110-7681
- Additional info contact
- Angela Drysdale 001-704-7768646
- Root cause description
- Nonconforming Material/Component
- Product quantity
- 860450
- Distribution pattern
- U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
- Product description
Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)
- Reason for recall
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
- Action
On 07/29/2026, an "URGENT MEDICAL DEVICE RECALL NOTICE" was emailed to customers informing them that, Greiner Bio-One North America identified a potential ethylenediaminetetraacetic acid (EDTA) contamination issue in a specific lot of VACUETTE Lithium Heparin serum separator tubes. Use of affected tubes may lead to: " Falsely elevated potassium (K+) results " Falsely decreased calcium (Ca++) results Customers are instructed to: " Immediately identify and quarantine all remaining units of item 454008P, lot B26043FM. " Stop using the affected lot for blood collection. " Notify relevant staff (phlebotomy, lab, clinical teams) of this advisory. " Review any recent patient results obtained using this lot. Consider repeat testing with a different lot if results appear inconsistent with clinical presentation. " Follow proper order of draw, and recollection procedures when re-testing. For unused product disposal (if destroyed by customer/distributor) " The unused blood collection tubes contain chemical additives (LH and serum separator). Recommend following OSHA process, and/or your facility's environmental health and safety (EHS) policies and procedures for disposal. For questions or documentation (Acknowledgement and Receipt Form) to Recalls.US@gbo.com or please call 800-515-8112.
- K numbers
- K960857
- Harmonized identifiers (openFDA)
- Device name
- Tubes, Vials, Systems, Serum Separators, Blood Collection
- Medical specialty description
- Clinical Chemistry
- Regulation number
- 862.1675
- Device class
- 2
- K number
- K952516, K823836, K120490, K103041, K250907, K924361, K232881, K926253, K964698, K200027, K831081, K882614, K202446, K902442, K043143, K993787, K790485, K230080, K944083, K252344, K931367, K153309, K935564, K162233, K030394, K823176, K012221, K991088, K991120, K945454, K900309, K123987, K981035, K252040, K911638, K791303, K940905, K101502, K950432, K833475 +200 more
- Registration number
- 3014570560, 3009976527, 1531062, 3014579161, 1125230, 3030896267, 3005941719, 2032112, 3013298431, 3031789039, 2025816, 3013188547, 3006950086, 2618282, 1219611, 9681572, 3012536737, 3037918935, 3016844398, 1832216, 3008449424, 3012187973, 3024537844, 3031233007, 3005587132, 1319639, 3014662855, 3007490412, 3010512490, 3009518015, 3030419987, 3014263632, 8010026, 3012809732, 3008337059, 1055927, 8020040, 3013696225, 3044466317, 3014312726 +181 more
- FEI number
- 3008002452, 3014570560, 3003737899, 3009976527, 3014579161, 3030896267, 3013298431, 3031789039, 3003560965, 2025816, 3002808075, 3002836339, 3013188547, 3006950086, 1219611, 3012536737, 3037918935, 3016844398, 1832216, 3008449424, 3012187973, 3024537844, 3031233007, 3005587132, 3014662855, 3007490412, 3010512490, 3009518015, 3030419987, 3014263632, 3012809732, 3008337059, 3013696225, 3044466317, 3014312726, 3012552122, 3010515547, 2828, 3043226252, 3002806476 +181 more