Device recalls

Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood C…

Greiner Bio-One North America, Inc. · Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) a…

Open, ClassifiedNonconforming Material/ComponentZ-0200-2027
Open FDA's page for this record
Cfres ID
221731
Product res number
Z-0200-2027
Event date initiated
2026-07-29
Event date posted
2026-10-21
Recall status
Open, Classified
Res event number
99518
Product code
JKA
Code info
Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12
Firm FEI number
3001451379
Recalling firm
Greiner Bio-One North America, Inc.
Address 1
4238 Capital Dr
City
Monroe
State
NC
Postal code
28110-7681
Additional info contact
Angela Drysdale 001-704-7768646
Root cause description
Nonconforming Material/Component
Product quantity
860450
Distribution pattern
U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
Product description

Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)

Reason for recall

Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.

Action

On 07/29/2026, an "URGENT MEDICAL DEVICE RECALL NOTICE" was emailed to customers informing them that, Greiner Bio-One North America identified a potential ethylenediaminetetraacetic acid (EDTA) contamination issue in a specific lot of VACUETTE Lithium Heparin serum separator tubes. Use of affected tubes may lead to: " Falsely elevated potassium (K+) results " Falsely decreased calcium (Ca++) results Customers are instructed to: " Immediately identify and quarantine all remaining units of item 454008P, lot B26043FM. " Stop using the affected lot for blood collection. " Notify relevant staff (phlebotomy, lab, clinical teams) of this advisory. " Review any recent patient results obtained using this lot. Consider repeat testing with a different lot if results appear inconsistent with clinical presentation. " Follow proper order of draw, and recollection procedures when re-testing. For unused product disposal (if destroyed by customer/distributor) " The unused blood collection tubes contain chemical additives (LH and serum separator). Recommend following OSHA process, and/or your facility's environmental health and safety (EHS) policies and procedures for disposal. For questions or documentation (Acknowledgement and Receipt Form) to Recalls.US@gbo.com or please call 800-515-8112.

K numbers
K960857
Harmonized identifiers (openFDA)
Device name
Tubes, Vials, Systems, Serum Separators, Blood Collection
Medical specialty description
Clinical Chemistry
Regulation number
862.1675
Device class
2
K number
K952516, K823836, K120490, K103041, K250907, K924361, K232881, K926253, K964698, K200027, K831081, K882614, K202446, K902442, K043143, K993787, K790485, K230080, K944083, K252344, K931367, K153309, K935564, K162233, K030394, K823176, K012221, K991088, K991120, K945454, K900309, K123987, K981035, K252040, K911638, K791303, K940905, K101502, K950432, K833475 +200 more
Registration number
3014570560, 3009976527, 1531062, 3014579161, 1125230, 3030896267, 3005941719, 2032112, 3013298431, 3031789039, 2025816, 3013188547, 3006950086, 2618282, 1219611, 9681572, 3012536737, 3037918935, 3016844398, 1832216, 3008449424, 3012187973, 3024537844, 3031233007, 3005587132, 1319639, 3014662855, 3007490412, 3010512490, 3009518015, 3030419987, 3014263632, 8010026, 3012809732, 3008337059, 1055927, 8020040, 3013696225, 3044466317, 3014312726 +181 more
FEI number
3008002452, 3014570560, 3003737899, 3009976527, 3014579161, 3030896267, 3013298431, 3031789039, 3003560965, 2025816, 3002808075, 3002836339, 3013188547, 3006950086, 1219611, 3012536737, 3037918935, 3016844398, 1832216, 3008449424, 3012187973, 3024537844, 3031233007, 3005587132, 3014662855, 3007490412, 3010512490, 3009518015, 3030419987, 3014263632, 3012809732, 3008337059, 3013696225, 3044466317, 3014312726, 3012552122, 3010515547, 2828, 3043226252, 3002806476 +181 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.