Device adverse event reports (MAUDE)

MiniMed 780G OUS System BLE Connect 3.0 mmol/L · AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL

Medtronic Puerto Rico Operations Co. · Key or Button Unresponsive/not Working

MalfunctionInitial submissionMDR 51243988
Manufacturer contact zip ext
-386
Manufacturer link flag
Y
Manufacturer g1 city
Juncos
Manufacturer contact address 1
Ceiba Norte Ind. Park #50 Road
Event type
Malfunction
Report number
2032227-2026-264260
Product problem flag
Y
Date received
20261030
PMA pmn number
Exempt
Date of event
20261021
Reprocessed and reused flag
N
Manufacturer contact zip code
00777
Reporter occupation code
003
Noe summarized
1
Manufacturer contact l name
Patel
Manufacturer g1 postal code
00777-3869
Reporter country code
UK
Date added
20261030
Manufacturer contact f name
Keya
Device date of manufacturer
20220706
Mfr report type
Thirty-Day
Manufacturer contact country
PR
Date changed
20261030
Health professional
N
Summary report flag
N
Manufacturer g1 zip code ext
-386
Date report
20261029
Initial report to FDA
No
Manufacturer g1 country
PR
Manufacturer contact city
Juncos
MDR report key
51243988
Date manufacturer received
20261021
Report source code
Manufacturer report
Manufacturer g1 zip code
00777
Manufacturer g1 name
Medtronic Puerto Rico Operations Co.
Adverse event flag
N
Single use flag
N
Manufacturer g1 address 1
Ceiba Norte Ind. Park #50 Road
Manufacturer contact postal code
00777-3869
Type of report
Initial submission
Source type
Foreign, Consumer
Device
  1. Device sequence number
    1821257
    Date received
    20261030
    Brand name
    MiniMed 780G OUS System BLE Connect 3.0 mmol/L
    Generic name
    AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL
    Manufacturer d name
    Medtronic Puerto Rico Operations Co.
    Manufacturer d address 1
    Ceiba Norte Ind. Park #50 Road
    Manufacturer d city
    Juncos
    Manufacturer d zip code
    00777
    Manufacturer d zip code ext
    -386
    Manufacturer d country
    PR
    Manufacturer d postal code
    00777-3869
    Device operator
    LAY USER/PATIENT
    Model number
    MMT-1885
    Catalog number
    MMT-1885
    Lot number
    HG6BUBKZZ
    Device availability
    No
    Device report product code
    OZP
    Combination product flag
    N
    UDI di
    000000763000639266
    UDI public
    (01)000000763000639266
    Harmonized identifiers (openFDA)
    Device name
    Automated Insulin Dosing Device System, Single Hormonal Control
    Medical specialty description
    Unknown
    Device class
    3
Product problems
Key or Button Unresponsive/not Working
Patient
  1. Patient sequence number
    1
    Patient problems
    No Clinical Signs, Symptoms or Conditions
MDR text
  1. MDR text key
    16629026500
    Text type code
    Additional Manufacturer Narrative
    Patient sequence number
    0
    Text

    Select patient information cannot be provided due to regional privacy regulations.\\The reported device is not marketed in the United States, but it is a same device to one that is marketed inside the United States.\\This MDR related to the Puerto Rico manufacturing site has been assigned a MedWatch number from the MiniMed Northridge site, per Variance 5.\\Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned. No conclusion can be drawn at this time. We therefore consider this report complete to the best of our knowledge.\\MiniMed submits this report to comply with FDA regulations 21 CFR Parts 4 and 803. MiniMed has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report. This report does not constitute an admission or a conclusion by FDA, MiniMed , or its employees that the device, MiniMed , or its employees caused or contributed to the event described in the report. Any required fields that are unpopulated are blank because the information is currently unknown or unavailable. MiniMed will submit a supplemental report if additional relevant information becomes known.

  2. MDR text key
    16129026499
    Text type code
    Description of Event or Problem
    Patient sequence number
    0
    Text

    It was reported to MiniMed that the customer experienced a Stuck Button alarm after airplane travel. The customer reported no adverse event. The event involved product(s) MMT-1885. Troubleshooting was performed, and the Stuck Button alarm was explained; the customer reported they had not pressed a button for 3 minutes or longer, was advised that atmospheric pressure changes may rarely cause the keypad to become unresponsive for up to 45 minutes, removed and reinstalled the battery cap without removing the battery, was able to clear the alarm and restore keypad response, did not receive a Battery Failed alarm, and was instructed to replace the battery as soon as possible, monitor the pump, and call back if issues persist. No harm requiring medical intervention was reported. No product return is required for MMT-1885.

Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

How to read this: Each report says someone told FDA an event happened while a product was used. A report does not show the product caused it, and counts rise and fall with how many people report. Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.