Device adverse event reports (MAUDE)

t:slim X2 Insulin Pump with Control-IQ+ Technology · Alternate Controller Enabled Insulin Infusion Pump

Tandem Diabetes Care · Obstruction of Flow

InjuryInitial submissionMDR 51243983
Manufacturer contact state
CA
Manufacturer link flag
Y
Manufacturer contact address 1
12400 High Bluff Drive
Manufacturer contact pcity
+1(858)2
Event type
Injury
Report number
3013756811-2026-197917
Product problem flag
Y
Date received
20261030
PMA pmn number
K250792
Date of event
20261009
Reprocessed and reused flag
N
Manufacturer contact zip code
92130
Manufacturer contact plocal
+1(858)224-6281
Noe summarized
1
Manufacturer contact l name
Connell
Reporter country code
US
Date added
20261030
Manufacturer contact f name
John
Device date of manufacturer
20251124
Previous use code
R
Reporter state code
OH
Mfr report type
Thirty-Day
Manufacturer contact country
US
Date changed
20261030
Health professional
N
Summary report flag
N
Date report
20261029
Manufacturer contact city
San Diego
MDR report key
51243983
Date manufacturer received
20261010
Report source code
Manufacturer report
Adverse event flag
Y
Single use flag
N
Manufacturer contact postal code
92130
Type of report
Initial submission
Source type
Consumer
Device
  1. Device sequence number
    1821253
    Date received
    20261030
    Brand name
    t:slim X2 Insulin Pump with Control-IQ+ Technology
    Generic name
    Alternate Controller Enabled Insulin Infusion Pump
    Manufacturer d name
    Tandem Diabetes Care
    Manufacturer d address 1
    12400 High Bluff Drive
    Manufacturer d city
    San Diego
    Manufacturer d state
    CA
    Manufacturer d zip code
    92130
    Manufacturer d country
    US
    Manufacturer d postal code
    92130
    Device operator
    LAY USER/PATIENT
    Model number
    1000354
    Catalog number
    1014073
    Device availability
    No
    Device report product code
    QFG
    Device evaluated by manufacturer
    N
    Combination product flag
    N
    UDI di
    00389152407319
    UDI public
    (01)00389152407319(22)1017690(11)251124
    Harmonized identifiers (openFDA)
    Device name
    Alternate Controller Enabled Insulin Infusion Pump
    Medical specialty description
    Clinical Chemistry
    Regulation number
    880.5730
    Device class
    2
Product problems
Obstruction of Flow
Patient
  1. Patient sequence number
    1
    Date received
    20261030
    Patient age
    20 YR
    Patient sex
    Unknown
    Patient weight
    135 LB
    Sequence number treatment
    Humalog - Insulin, Tandem - Cartridge, Unomedical - Infusion Set
    Sequence number outcome
    Hospitalization, R
    Patient problems
    Hyperglycemia, Elevated Ketones/ Diabetic Ketoacidosis
MDR text
  1. MDR text key
    16128963597
    Text type code
    Description of Event or Problem
    Patient sequence number
    0
    Text

    It was reported that pump issued occlusion alarms resulting in the customer going to the emergency room the next date, (b)(6) 2026. The customer was admitted to the hospital's intensive care unit with a blood glucose level of 426 mg/dL, a healthcare provider identified ketone level as dangerous or life threatening. Customer was treated with intravenous fluids of saline and insulin which resolved the issue. The customer was released the same day with issue resolved and no permanent damage. Customer reported using the same cartridge for more than 48 hours and received education from Technical Support to change cartridge every 48 hours with Humalog brand insulin, after which case was closed.

  2. MDR text key
    16628963598
    Text type code
    Additional Manufacturer Narrative
    Patient sequence number
    0
    Text
    In use at the time of the event:\\CGM model: Unknown.\\Mobile OS: iOS / iOS_iPhone XS / Unknown / iOS_17.3.1.
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

How to read this: Each report says someone told FDA an event happened while a product was used. A report does not show the product caused it, and counts rise and fall with how many people report. Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.