Device recall enforcement reports

Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695

Baxter Healthcare Corporation · There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, th…

Class IIOngoingZ-0351-2027
Status
Ongoing
City
Deerfield
State
IL
Country
United States
Classification
Class II
Product type
Devices
Event ID
100117
Recalling firm
Baxter Healthcare Corporation
Address 1
1 Baxter Pkwy
Postal code
60015-4625
Voluntary mandated
Voluntary: Firm initiated
Initial firm notification
Letter
Distribution pattern
Worldwide - US Nationwide distribution.
Recall number
Z-0351-2027
Product description
Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695
Product quantity
77 units
Recall initiation date
20260929
Center classification date
20261103
Report date
20261111
Reason for recall

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.

Code info

Serial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411, 108516637, 108544106, 108544116, 108618964, 108463224, 108484405, 108489648, 108511412, 108521436, 108544107, 108604137, 108618965, 108468062, 108484406, 108506316, 108511413, 108521452, 108544110, 108604140, 108468063, 108484408, 108506317, 108516634, 108536776, 108544111, 108604153; UDI-DI number: 00887761974425