Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlea…

Cochlear Limited · Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits

Class IIOngoingZ-0349-2027
Status
Ongoing
City
Macquarie University
Country
Australia
Classification
Class II
Product type
Devices
Event ID
100098
Recalling firm
Cochlear Limited
Address 1
1 University Ave
Postal code
N/A
Voluntary mandated
Voluntary: Firm initiated
Initial firm notification
Letter
Distribution pattern
U.S.: NJ, NY, and PA. O.U.S.: Not provided
Recall number
Z-0349-2027
Product quantity
4 device in the U.S.
Reason for recall
Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits
Recall initiation date
20260930
Center classification date
20261103
Report date
20261111
Code info
Model: CI1000 Product/Catalog: P1979576 UDI-DI code: 9321502069847 Serial Numbers: 1020210051972 1020210070461 1020210145046 1020210048286
Product description

Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, single use devices intended for long term implantation under the skin in the mastoid region on either side of the head. The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work.