Device recall enforcement reports

TruSystem 7000 U Operating Table (dV), Product Code 1723633

Baxter Healthcare Corporation · The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occ…

Class IIOngoingZ-0332-2027
Status
Ongoing
City
Round Lake
State
IL
Country
United States
Classification
Class II
Product type
Devices
Event ID
99826
Recalling firm
Baxter Healthcare Corporation
Address 1
25212 W Il Route 120
Postal code
60073-9799
Voluntary mandated
Voluntary: Firm initiated
Initial firm notification
Letter
Distribution pattern
US Nationwide distribution in the states of IL, MO, TX.
Recall number
Z-0332-2027
Product description
TruSystem 7000 U Operating Table (dV), Product Code 1723633
Product quantity
62 units
Recall initiation date
20260914
Center classification date
20261103
Report date
20261111
Code info
UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662
Reason for recall

The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.