Senographe Pristina

GE Medical Systems, LLC · GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisitio…

Class IIOngoingZ-0327-2027
Status
Ongoing
City
Waukesha
State
WI
Country
United States
Classification
Class II
Product type
Devices
Event ID
100070
Recalling firm
GE Medical Systems, LLC
Address 1
3000 N Grandview Blvd
Postal code
53188-1615
Voluntary mandated
FDA Mandated
Distribution pattern
USA and OUS.
Recall number
Z-0327-2027
Product description
Senographe Pristina
Recall initiation date
20260828
Center classification date
20261104
Report date
20261111
Reason for recall

GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).

Code info

GTIN(s): 00840682118460, 00840682142052, 00840682145879, 00195278022745, 00195278276971, 00195278661999, 00195278665379, 00195278914026, 00198953109281, 10198953054311, 10195278901764.