Senographe Pristina
GE Medical Systems, LLC · GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisitio…
- Status
- Ongoing
- City
- Waukesha
- State
- WI
- Country
- United States
- Classification
- Class II
- Product type
- Devices
- Event ID
- 100070
- Recalling firm
- GE Medical Systems, LLC
- Address 1
- 3000 N Grandview Blvd
- Postal code
- 53188-1615
- Voluntary mandated
- FDA Mandated
- Distribution pattern
- USA and OUS.
- Recall number
- Z-0327-2027
- Product description
- Senographe Pristina
- Recall initiation date
- 20260828
- Center classification date
- 20261104
- Report date
- 20261111
- Reason for recall
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
- Code info
GTIN(s): 00840682118460, 00840682142052, 00840682145879, 00195278022745, 00195278276971, 00195278661999, 00195278665379, 00195278914026, 00198953109281, 10198953054311, 10195278901764.